UNITED
STATES ENVIRONMENTAL PROTECTION AGENCY
NATIONAL EXPOSURE RESEARCH LABORATORY
HUMAN EXPOSURE & ATMOSPHERIC SCIENCES DIVISION (MD-D205-03)
Research Triangle Park, NC 27711
919-541-3737
Office of Research and Development
LIST OF DESIGNATED REFERENCE AND EQUIVALENT METHODS
Issue Date:
August 16, 2004
(www.epa.gov/ttn/amtic/criteria.html)
These methods for measuring ambient concentrations of specified air
pollutants have been designated as "reference methods" or "equivalent
methods" in accordance with Title 40, Part 53 of the Code of Federal
Regulations (40 CFR Part 53). Subject to any limitations (e.g.,
operating range or temperature range) specified in the applicable
designation, each method is acceptable for
use in state or local air quality surveillance systems under 40 CFR Part
58 unless the applicable designation is subsequently canceled. Automated
methods for pollutants other than PM10
are acceptable for use only at shelter temperatures between 20°C and
30°C and line voltages between 105 and 125 volts unless wider limits are
specified in the method description.
Prospective users of the methods listed should note (1) that each method
must be used in strict accordance with its associated operation or
instruction manual and with applicable quality assurance procedures, and
(2) that modification of a method by its vendor or user may cause the
pertinent designation to be inapplicable to the method as modified. (See
Section 2.8 of Appendix
C, 40 CFR Part 58 for approval of modifications to any of these methods
by users.)
Further information concerning particular designations may be found in
the
Federal Register
notice cited for each method or by writing to the National Exposure
Research Laboratory, Human Exposure and Atmospheric Sciences Division
(MD-D205-03), U.S. Environmental Protection Agency, Research Triangle
Park, North Carolina 27711. Technical information concerning the methods
should be obtained by contacting the source listed for each method.
Source addresses are listed at the end of the listing of methods, except
for the addresses for lead method sources, which are given with the
method. New analyzers or PM10
samplers
sold as reference or equivalent methods must carry a label or sticker
identifying them as designated methods. For analyzers or PM10
or samplers sold prior to the designation of a method with the same or
similar model number, the model number does not necessarily identify an
analyzer or sampler as a designated method. Consult the manufacturer or
seller to determine if a previously
sold analyzer or sampler can be considered a designated method or if it
can be upgraded to designation status. Analyzer users who experience
operational or other difficulties with a designated analyzer or sampler
and are unable to resolve the problem directly with the instrument
manufacturer may contact EPA (preferably in writing) at the above
address for assistance.
This list will be revised as necessary to reflect any new designations
or any cancellation of a designation currently in effect.
The most current revision of the list will be available for inspection
at EPA's Regional Offices, and copies may be obtained at the Internet
site identified above or by writing to the National Exposure Research
Laboratory at the address specified above.
Most Recent Designations
SIR S.A. Model S-5012 NO/NOx/NO2
Analyzer Aug. 06, 2004
Thermo Electron Models RAAS2.5-100 FEM, RAAS2.5-200 FEM, and
RAAS2.5-300 FEM VSCC PM2.5
samplers Aug. 06, 2004
Environnement S.A. Model MP101M PM10
Beta Gauge Monitor April 08, 2004
TSP...PM10
PARTICULATE MATTER - TSP
Reference Method for TSP
Manual Reference Method:
40 CFR Part 50, Appendix B
Reference Method for the Determination of Suspended Particulate Matter
in the Atmosphere (High-Volume Method)
[Federal Register: Vol 47, page 54912, 12/06/82 and Vol 48, page 17355,
04/22/83]
PARTICULATE MATTER - PM10
Andersen Model RAAS10-100 PM10 Single Channel PM10
Sampler
Manual Reference Method:
RFPS-0699-130
“Andersen Instruments, Incorporated Model RAAS10-100 Single Channel
Reference Method PM10
Sampler,” with RAAS-10 PM10
inlet or the louvered inlet specified in 40 CFR 50 Appendix L, Figs. L-2
thru L-19, configured as a PM10
reference method, and operated
for 24-hour continuous sample periods at a flow rate of 16.67 liters/
minute, and in accordance with the Model RAAS105-100 Operator’s
Manual and with the requirements and sample collection filters specified
in 40 CFR Part 50, Appendix J or Appendix M.
[Federal
Register:
Vol 64, page 33481, 06/23/99]
Andersen Model RAAS10-200 PM10 Single Channel PM10
Audit Sampler
Manual Reference Method:
RFPS-0699-131
“Andersen Instruments, Incorporated Model RAAS10-200 Single Channel
Reference Method PM10
Audit Sampler,” with RAAS-10
PM10
inlet or the louvered inlet specified in 40 CFR 50 Appendix L, Figs. L-2
thru L-19, configured as a PM10
reference method, and
operated for 24-hour continuous sample periods at a flow rate of 16.67
liters/minute, and in accordance with the Model RAAS105-200
Operator’s Manual and with the requirements and sample collection
filters specified in 40 CFR Part 50, Appendix J or Appendix M.
[Federal
Register:
Vol 64, page 33481, 06/23/99]
Andersen Model RAAS10-300 PM10 Multi Channel PM10
Sampler
Manual Reference Method:
RFPS-0699-132
“Andersen Instruments, Incorporated Model RAAS10-300 Multi Channel
Sequential Reference Method PM10
Sampler,” with RAAS-10
PM10
inlet or the louvered inlet specified in 40 CFR 50 Appendix L, Figs. L-2
thru L-19, configured as a PM10
reference method, and
operated for 24-hour continuous sample periods at a flow rate of 16.67
liters/minute, and in accordance with the Model RAAS105-300
Operator’s Manual and with the requirements and sample collection
filters specified in 40 CFR Part 50, Appendix J or Appendix M.
[Federal
Register:
Vol 64, page 33481, 06/23/99]
BGI Incorporated Model PQ100 Air Sampler
Manual Reference Method:
RFPS-1298-124
“BGI Incorporated Model PQ100 Air Sampler” with BGI 16.7 Inlet Kit or
the louvered inlet specified in 40 CFR 50 Appendix L, Figs.
L-2 thru L-19, configured as a PM10
reference method, for 24-hour continuous sample periods at a flow rate
of 16.7 liters/minute,
operated in accordance with the Model PQ100 Instruction Manual and with
the requirements specified in 40 CFR Part 50, Appendix
J or Appendix M, using either the original or the newer PQ200-type
filter cassettes, and with or without the optional Solar Panel Power
Supply. [Federal
Register:
Vol 63, page 69625, 12/17/98]
BGI Incorporated Model PQ200 Air Sampler
Manual Reference Method:
RFPS-1298-125
“BGI Incorporated Model PQ200 Air Sampler” with “flat plate” PM10
inlet or the louvered inlet specified in 40 CFR 50 Appendix L,
Figs. L-2 thru L-19, configured as a PM10
reference method, and operated for 24-hour continuous sample periods in
accordance with
the Model PQ200 Instruction Manual and with the requirements specified
in 40 CFR Part 50, Appendix J or Appendix M, and with or
without the optional Solar Panel Power Supply. [Federal
Register:
Vol 63, page 69625, 12/17/98]
Environnement S.A. Model MP101M PM10
Monitor
Automated Equivalent Method:
EQPM-0404-151
“Environnement
S. A. Model MP101M PM10
Beta Gauge Monitor,” configured with the louvered PM10
inlet specified in 40 CFR 50
Appendix L or its flat-topped predecessor version and one of the three
optional temperature-regulated sampling tubes (RST), and
operated with a full scale measurement range of
0 - 0.500 mg/m3
(0 - 500
:g/m3),
with the sample flow rate set to
1.00 m3/h
and flow
regulation set to
yes,
the “norms selection” set to
m3
(actual volume), the “cycle” set to
24 hours,
the “period” set to
none,
and the
“counting time” set to
200 seconds.2
[Federal
Register:
Vol. 69, page 18569, 4/8/04]
PM10
Graseby Andersen/GMW Model 1200 High-Volume Air Sampler
Manual Reference Method:
RFPS-1287-063
Sierra-Andersen or General Metal Works Model 1200 PM10
High-Volume Air Sampler System," consisting of a Sierra-Andersen or
General Metal Works Model 1200 PM10
Size-Selective Inlet and any of the high-volume air samplers identified
as SAUV-10H,
SAUV-11H, GMW-IP-10, GMW-IP-10-70, GMW-IP-10-801, or GMW-IP-10-8000,
which include the following components: Anodized
aluminum high-volume shelter with either acrylonitrile butadiene styrene
plastic filter holder and motor/blower housing or stainless steel
filter holder and phenolic plastic motor/blower housing; 0.6 hp
motor/blower; pressure transducer flow recorder; either an electronic
mass flow controller or a volumetric flow controller; either a digital
timer/programmer, seven-day mechanical timer, six-day
timer/programmer, or solid-state timer/programmer; elapsed time
indicator; and filter cartridge.
[Federal
Register:
Vol 52, page 45684, 12/01/87 and Vol 53, page 1062, 01/15/88]
Graseby Andersen/GMW Model 321-B High-Volume Air Sampler
Manual Reference Method:
RFPS-1287-064
"Sierra-Andersen or General Metal Works Model 321-B PM10
High-Volume Air Sampler System," consisting of a Sierra-Andersen or
General Metal Works Model 321-B PM10
Size-Selective Inlet and any of the high-volume air samplers identified
as SAUV-10H,
SAUV-11H, GMW-IP-10, GMW-IP-10-70, GMW-IP-10-801, or GMW-IP-10-8000,
which include the following components:
Anodized aluminum high-volume shelter with either acrylonitrile
butadiene styrene plastic filter holder and motor/blower housing or
stainless steel filter holder and phenolic plastic motor/blower housing;
0.6 hp motor/blower; pressure transducer flow recorder; either
an electronic mass flow controller or a volumetric flow controller;
either a digital timer/programmer, seven-day mechanical timer, six-day
timer/programmer, or solid-state timer/programmer; elapsed time
indicator; and filter cartridge.
[Federal
Register:
Vol 52, page 45684, 12/01/87 and Vol 53, page 1062, 01/15/88]
Graseby Andersen/GMW Model 321-C High-Volume Air Sampler
Manual Reference Method:
RFPS-1287-065
"Sierra-Andersen or General Metal Works Model 321-C PM10
High-Volume Air Sampler System," consisting of a Sierra-Andersen
General Metal Works Model 321-C PM10
or Size-Selective Inlet and any of the high-volume air samplers
identified as SAUV-10H,
SAUV-11H, GMW-IP-10, GMW-IP-10-70, GMW-IP-10-801, or GMW-IP-10-8000,
which include the following components:
Anodized aluminum high-volume shelter with either acrylonitrile
butadiene styrene plastic filter holder and motor/blower housing or
stainless steel filter holder and phenolic plastic motor/blower housing;
0.6 hp motor/blower; pressure transducer flow recorder; either
an electronic mass flow controller or a volumetric flow controller;
either a digital timer/programmer, seven-day mechanical timer, six-day
timer/programmer, or solid-state timer/programmer; elapsed time
indicator; and filter cartridge.
[Federal
Register:
Vol 52, page 45684, 12/01/87 and Vol 53, page 1062, 01/15/88]
Graseby Andersen/GMW Models SA241 and SA241M Dichotomous Sampler
Manual Reference Method:
RFPS-0789-073
"Sierra-Andersen Models SA241 and SA241M or General Metal Works Models
G241 and G241M PM10
Dichotomous Samplers,"
consisting of the following components: Sampling Module with SA246b or
G246b 10 µm inlet or the louvered inlet specified in 40 CFR
50 Appendix L, Figs. L-2 thru L-19, 2.5 µm virtual impactor assembly, 37
mm coarse and fine particulate filter holders, and tripod
mount; Control Module with diaphragm vacuum pump, pneumatic constant
flow controller, total and coarse flow rotameters and
vacuum gauges, pressure switch (optional), 24-hour flow/event recorder,
digital timer/programmer or 7-day skip timer, and elapsed time
indicator. [Federal
Register:
Vol 54, page 31247, 07/27/89]
Graseby Andersen/GMW Model FH621-N Beta Monitor
Automated Equivalent Method:
EQPM-0990-076
"Andersen Instruments Model FH62I-N PM10
Beta Attenuation Monitor," consisting of the following components: FH101
Vacuum
Pump Assembly; FH102 Accessory Kit; FH107 Roof Flange Kit; FH125 Zero
and Span PM10
Mass Foil Calibration Kit; FH62I Beta
Attenuation 19-inch Control Module; SA246b PM10
Inlet (16.7 liter/min) or the louvered inlet specified in 40 CFR 50
Appendix L, Figs.
L-2 thru L-19; operated for 24-hour average measurements, with an
observing time of 60 minutes, the calibration factor set to 2400, a
glass fiber filter tape, an automatic filter advance after each 24-hour
sample period, and with or without either of the following options:
FH0P1 Indoor Cabinet; FH0P2 Outdoor Shelter Assembly. [Federal
Register:
Vol 55, page 38387, 09/18/90]
Met One or Sibata Models BAM/GBAM 1020, BAM/GBAM 1020-1
Automated Equivalent Method:
EQPM-0798-122
“Met One Instruments or Sibata Scientific Technology Models BAM 1020,
GBAM 1020, BAM 1020-1, and GBAM 1020-1 PM10 Beta
Attenuation Monitor,” including the BX-802 sampling inlet, operated for
24-hour average measurements, with a filter change frequency
of one hour, with glass fiber filter tape, and with or without any of
the following options: BX-823, tube extension; BX-825, heater kit;
BX-826, 230 Vac heater kit; BX-828, roof tripod; BX-902, exterior
enclosure; BX-903, exterior enclosure with temperature control; BX-
961, mass flow controller; BX-967, internal calibration device. [Federal
Register:
Vol 63, page 41253,08/03/98]
Oregon DEQ Medium Volume PM10
Sampler
Manual Reference Method:
RFPS-0389-071
"Oregon DEQ Medium Volume PM10
Sampler." NOTE: This method is not now commercially available.
[Federal
Register:
Vol 54, page 12273,03/24/89]
PM10
Rupprecht & Patashnick TEOM Series 1400/1400a PM10
Monitors
Automated Equivalent Method:
EQPM-1090-079
"Rupprecht & Patashnick TEOM Series 1400 and Series 1400a PM-10
Monitors" (including serial number prefixes 1400, 140A, 140AA,
140AB, 140AT, and 140UP), consisting of the following components: TEOM
Sensor Unit; TEOM Control Unit; Flow Splitter (3
liter/min sample flow); Teflon-Coated Glass Fiber Filter Cartridges;
Rupprecht & Patashnick PM-10 Inlet (part number 57-00596),
Sierra-Andersen Model 246b PM-10 Inlet (16.7 liter/min) or louvered
inlet specified in 40 CFR 50 Appendix L, Figs. L-2 thru L-19;
operated for 24-hour average measurements, with the total mass averaging
time set at 300 seconds, the mass rate/mass concentration
averaging time set at 300 seconds, the gate time set at 2 seconds, and
with or without any of the following options: Tripod; Outdoor
Enclosure; Automatic Cartridge Collection Unit (Series 1400a only); Flow
Splitter Adapter (for 1 or 2 liter/min sample flow).
[Federal
Register:
Vol 55, page 43406, 10/29/90]
Rupprecht & Patashnick Partisol Model 2000 Air Sampler
Manual Reference Method:
RFPS-0694-098
"Rupprecht & Patashnick Partisol Model 2000 Air Sampler," consisting of
a Hub Unit and 0, 1, 2, or 3 Satellite Units, with each
sampling station used for PM10
measurements equipped with a Rupprecht & Patashnick PM-10 inlet and
operated for continuous 24-hour
periods using the Basic, Manual, Time, Analog Input, or Serial Input
programming modes, and with or without any of the following
options: PM2.5-
style filter cassette holder; louvered inlet specified in 40 CFR 50
Appendix L, Figs. L-2 thru L-19 in lieu of standard inlet;
57-002320 Stand for Hub or Satellite; 59-002542 Advanced EPROM;
10-001403 Large Pump (1/4 hp); 120 VAC. Hardware for Indoor
Installation consists of: 51-002638-xxxx Temperature Sensor (Extended
Length); 55-001289 Roof Flange (1 1/4"); 57-000604 Support
Tripod for Inlet; 57-002526-0001 Sample Tube Extension (1 m);
57-002526-0002 Sample Tube Extension (2 m). Hardware for Outdoor
Installation in Extreme Cold Environments consists of: 10-002645
Insulating Jacket for Hub Unit.
[Federal
Register:
Vol 59, page 35338, 07/11/94]
Rupprecht and Patashnick Co. Partisol®-FRM Model 2000 PM10
Air Sampler
Manual Reference Method:
RFPS-1298-126
“Rupprecht and Patashnick Company Partisol®-FRM Model 2000 PM10
Air Sampler” with PM10 inlet or louvered inlet specified in
40 CFR 50 Appendix L, Figs. L-2 thru L-19, configured as a PM10
reference method, and operated for 24-hour continuous sample periods
in accordance with the Model 2000 Instruction Manual and with the
requirements specified in 40 CFR Part 50, Appendix J or Appendix
M. [Federal
Register:
Vol 63, page 69625, 12/17/98]
Rupprecht and Patashnick Partisol®-Plus Model 2025 PM10
Seq. Air Sampler
Manual Reference Method:
RFPS-1298-127
“Rupprecht and Patashnick Company Partisol®-Plus Model 2025 PM10
Sequential Air Sampler” with PM10
inlet or louvered inlet
specified in 40 CFR 50 Appendix L, Figs. L-2 thru L-19, configured as a
PM10
reference method, and operated for 24-hour continuous
sample periods in accordance with the Model 2025 Instruction Manual and
with the requirements specified in 40 CFR Part 50, Appendix
J or Appendix M. [Federal
Register:
Vol 63, page 69625, 12/17/98]
Thermo Andersen Series FH 62 C14 Continuous PM10 Monitor
Automated Equivalent Method:
EQPM-1102-150
“Thermo
Andersen Series FH 62 C14 Continuous PM10 Ambient Particulate Monitor,”
operated for 24-hour average measurements,
with the specified 10-micron inlet, inlet connector, sample tube with
heater, roof flange kit, mass foil kit, pump kit, sample filter tape;
with operational settings of 1000 L/h (16.67 L/min) sample flow rate,
daily filter change, auto filter change at volumetric flow < 950
L/h, auto filter change at mass > 1500 micrograms, and factory default
calculation mode settings; and with operational calibration and
servicing as outlined in the Operator Manual. [Federal
Register:
Vol 67, page 76174, 12/11/02]
Tisch Environmental Model TE-6070 PM10 High-Volume Air Sampler or
New Star Environmental Model NS-6070 PM10 High-Volume Air Sampler
Manual Reference Method:
RFPS-0202-141
“Tisch Environmental Model TE-6070 or New Star Environmental Model
NS-6070 PM10 High-Volume Air Sampler,” consisting of
a TE-6001 PM10 size-selective inlet, 8" x 10" filter holder, aluminum
outdoor shelter, mass flow controller or volumetric flow controller
with brush or brushless motor, 7-day mechanical off/on-elapsed timer or
11-day digital off/on-elapsed timer, and any of the high volume
sampler variants identified as TE-6070-BL or NS-6070-BL, TE-6070D or
NS-6070D, TE-6070D-BL or NS-6070-BL, TE-6070V or NS-
6070V, TE-6070V-BL or NS-6070V-BL, TE-6070-DV or NS-6070-DV, or
TE-6070DV-BL or NS-6070DV-BL, with or without the
optional stainless steel filter media holder/filter cartridge or
continuous flow/pressure recorder.
[Federal
Register:
Vol 67, page 15566, 04/02/02]
Wedding & Associates' or Thermo Environmental Instruments Inc.
Model 600 PM10
High-Volume Sampler
Manual Reference Method:
RFPS-1087-062
"Wedding & Associates' or Thermo Environmental Instruments, Inc. Model
600 PM10
Critical Flow High-Volume Sampler," consisting
of the following W&A/TEII components: PM10
Inlet; Critical Flow Device; Anodized Aluminum Shelter; Blower Motor
Assembly for
115, 220 or 240 VAC and 50/60 Hz; Mechanical Timer; Elapsed Time
Indicator; and Filter Cartridge/Cassette, and with or without the
following options: Digital Timer, 6 or 7 Day Timer, and 1 or 7 Day
Pressure Recorder.
[Federal
Register:
Vol 52, page 37366,10/06/87]
PM10...PM2.5
Wedding & Associates' or Thermo Environmental Instruments Inc.
Model 650 PM10
Beta Gauge
Automated Equivalent Method:
EQPM-0391-081
"Wedding & Associates' or Thermo Environmental Instruments, Inc. Model
650 PM10
Beta Gauge Automated Particle Sampler,"
consisting of the following W&A/TEII components: Particle Sampling
Module, PM10
Inlet (18.9 liter/min), Inlet Tube and Support
Ring, Vacuum Pump (115, 220 or 240 VAC and 50/60 Hz); and operated for
24-hour average measurements with glass fiber filter
tape. [Federal
Register:
Vol 56, page 9216, 03/05/91]
PARTICULATE MATTER - PM2.5
Andersen Model RAAS2.5-200 PM2.5 Ambient Audit Air Sampler
Manual Reference Method:
RFPS-0299-128
“Andersen Instruments, Incorporated Model RAAS2.5-200 PM2.5 Audit
Sampler,” configured as a PM2.5
reference method and operated
with software (firmware) version 4B, 5.0.1 - 6.09, 6.0A, or 6.0B, for
24-hour continuous sample periods at a flow rate of 16.67
liters/minute, and in accordance with the Model RAAS2.5-200 Operator’s
Manual and with the requirements and sample collection filters
specified in 40 CFR Part 50, Appendix L. [Federal
Register:
Vol 64, page 12167, 03/11/99]
BGI Inc. Models PQ200 or PQ200A PM2.5
Ambient Fine Particle Sampler
Manual Reference Method:
RFPS-0498-116
“BGI Incorporated Models PQ200 and PQ200A PM2.5
Ambient Fine Particle Sampler,” operated with firmware version 3.88 or
3.89R,
for 24-hour continuous sample periods, in accordance with the Model
PQ200/PQ200A Instruction Manual and with the requirements
and sample collection filters specified in 40 CFR Part 50, Appendix L,
and with or without the optional Solar Power Supply or the
optional dual-filter cassette (P/N F-21/6) and associated lower impactor
housing (P/N B2027), where the upper filter is used for PM2.5.
The Model PQ200A is described as a portable audit sampler and includes a
set of three carrying cases.
[Federal
Register:
Vol 63, page 18911, 04/16/98]
BGI Inc. Models PQ200-VSCC or PQ200A-VSCC PM2.5
Sampler
Manual Equivalent Method:
EQPM-0202-142
“BGI Incorporated Models PQ200-VSCC or PQ200A-VSCC PM2.5
Ambient Fine Particle Sampler,” configured with a BGI VSCC™
Very Sharp Cut Cyclone particle size separator (in lieu of a WINS
impactor) and operated with firmware version 3.88, 3.91, 3.89R, or
3.91R, for 24-hour continuous sample periods, in accordance with the
Model PQ200/PQ200A Instruction Manual and VSCC
supplemental manual and with the requirements and sample collection
filters specified in 40 CFR Part 50, Appendix L, and with or
without the optional Solar Power Supply or the optional dual-filter
cassette (P/N F-21/6) and associated lower impactor housing (P/N
B2027), where the upper filter is used for PM2.5.
The Model PQ200A VSCC is described as a portable audit sampler and
includes a set
of three carrying cases. [Federal
Register:
Vol 67, page 15567, 04/02/02]
Graseby Andersen Model RAAS2.5-100 PM2.5 Ambient Air Sampler
Manual Reference Method:
RFPS-0598-119
“Graseby Andersen Model RAAS2.5-100 PM2.5 Ambient Air Sampler,” operated
with software version 4B, 5.0.1 - 6.09, 6.0A, or 6.0B,
configured for “Single 2.5” operation, for 24-hour continuous sample
periods at a flow rate of 16.67 liters/minute, and in accordance
with the Model RAAS2.5-100 Operator’s Manual and with the requirements
and sample collection filters specified in 40 CFR Part 50,
Appendix L. [Federal
Register:
Vol 63, page 31991, 06/11/98]
Graseby Andersen Model RAAS2.5-300 PM2.5 Sequential Ambient Air Sampler
Manual Reference Method:
RFPS-0598-120
“Graseby Andersen Model RAAS2.5-300 PM2.5 Sequential Ambient Air
Sampler,” operated with software version 4B, 5.0.1 - 6.09,
6.0A, or 6.0B, configured for “Multi 2.5" operation, for 24-hour
continuous sample periods at a flow rate of 16.67 liters/minute, and in
accordance with the Model RAAS2.5-300 Operator’s Manual and with the
requirements and sample collection filters specified in 40
CFR Part 50, Appendix L. [Federal
Register:
Vol 63, page 31991, 06/11/98]
Rupprecht & Patashnick Partisol®-FRM
Model 2000 PM-2.5 Air Sampler
Manual Reference Method:
RFPS-0498-117
"Rupprecht & Patashnick Company, Incorporated Partisol®-FRM Model 2000
PM-2.5 Air Sampler," operated with software versions
1.102 - 1.202, with either R&P-specified machined or molded filter
cassettes, with or without the optional insulating jacket for cold
weather operation, for 24-hour continuous sample periods, in accordance
with the Model 2000 Instruction Manual and with the
requirements and sample collection filters specified in 40 CFR Part 50,
Appendix L.
[Federal
Register:
Vol 63, page 18911, 04/16/98]
PM2.5
Rupprecht & Patashnick Partisol®-FRM
Model 2000 PM-2.5 FEM Air Sampler
Manual Equivalent Method:
EQPM-0202-143
"Rupprecht & Patashnick Co., Inc. Partisol®-FRM
Model 2000 PM-2.5 FEM Air Sampler," configured with a BGI VSCC™ Very
Sharp
Cut Cyclone particle size separator (in lieu of a WINS impactor) and
operated with software versions 1.102 - 1.202, with either R&Pspecified
machined or molded filter cassettes, for 24-hour continuous sample
periods, in accordance with the Model 2000 Instruction
Manual and VSCC supplemental manual, with the requirements and sample
collection filters specified in 40 CFR Part 50, Appendix L,
and with or without the optional insulating jacket for cold weather
operation. [Federal
Register:
Vol 67, page 15567, 04/02/02]
Rupprecht & Patashnick Partisol®
Model 2000 PM-2.5 Audit Sampler
Manual Reference Method:
RFPS-0499-129
“Rupprecht & Patashnick Company, Inc. Partisol®
Model 2000 PM-2.5 Audit Sampler,” configured as a PM2.5
reference method and
operated with software (firmware) version 1.2 - 1.202, for 24-hour
continuous sample periods at a flow rate of 16.67 liters/minute, in
accordance with the Partisol®
Model 2000 Operating Manual and with the requirements and sample
collection filters specified in 40 CFR
Part 50, Appendix L. [Federal
Register:
Vol 64, page 19153, 04/19/99]
Rupprecht & Patashnick Partisol®
Model 2000 PM-2.5 FEM Audit Sampler
Manual Equivalent Method:
EQPM-0202-144
“Rupprecht & Patashnick Co., Inc. Partisol®
Model 2000 PM-2.5 FEM Audit Sampler,” configured with a BGI VSCC™ Very
Sharp
Cut Cyclone particle size separator (in lieu of a WINS impactor), and
operated with software (firmware) version 1.2 - 1.202, for 24-hour
continuous sample periods at a flow rate of 16.67 liters/minute, in
accordance with the Partisol®
Model 2000 Operating Manual and
VSCC supplemental manual and with the requirements and sample collection
filters specified in 40 CFR Part 50, Appendix L.
[Federal
Register:
Vol 67, page 15567, 04/02/02]
Rupprecht & Patashnick Partisol®-Plus
Model 2025 Sequential Air Sampler
Manual Reference Method:
RFPS-0498-118
"Rupprecht & Patashnick Company, Incorporated Partisol®-Plus Model 2025
PM-2.5 Sequential Air Sampler," operated with any
software version 1.003 through1.413, with either R&P-specified machined
or molded filter cassettes, for 24-hour continuous sample
periods, in accordance with the Model 2025 Instruction Manual and with
the requirements and sample collection filters specified in 40
CFR Part 50, Appendix L. [Federal
Register:
Vol 63, page 18911, 04/16/98]
Rupprecht & Patashnick Partisol®-Plus
Model 2025 FEM Sequential Sampler
Manual Equivalent Method:
EQPM-0202-145
"Rupprecht & Patashnick Co., Inc. Partisol®-Plus
Model 2025 PM-2.5 FEM Sequential Air Sampler," configured with a BGI
VSCC™
Very Sharp Cut Cyclone particle size separator (in lieu of a WINS
impactor), and operated with any software version 1.003 through
1.413, with either R&P-specified machined or molded filter cassettes,
for 24-hour continuous sample periods, in accordance with the
Model 2025 Instruction Manual and VSCC supplemental manual and with the
requirements and sample collection filters specified in
40 CFR Part 50, Appendix L. [Federal
Register:
Vol 67, page 15567, 04/02/02]
Thermo Electron Model RAAS2.5-100 FEM PM2.5
Ambient Air Sampler
Manual Equivalent Method:
EQPM-0804-153
“Thermo Electron Corporation Model RAAS2.5-100 FEM” PM2.5
Ambient Air Sampler, configured with a BGI VSCC™ Very Sharp
Cut Cyclone particle size separator and operated with software version
06.0B.00 configured for “Single 2.5” operation, for 24-hour
continuous sample periods at a flow rate of 16.67 liters/minute, in
accordance with the Model RAAS2.5-100 FEM Operator’s Manual
and VSCC™ supplemental manual, and in accordance with the requirements
and sample collection filters specified in 40 CFR Part 50,
Appendix L [Federal
Register:
Vol 69, page 47924, 08/06/04]
Thermo Electron Model RAAS2.5-200 FEM PM2.5
Audit Air Sampler
Manual Equivalent Method:
EQPM-0804-154
“Thermo Electron Corporation Model RAAS2.5-200 FEM” PM2.5
Audit Air Sampler, configured with a BGI VSCC™ Very Sharp Cut
Cyclone particle size separator and operated with software version
06.0B.00, for 24-hour continuous sample periods at a flow rate of
16.67 liters/minute, in accordance with the Model RAAS2.5-200 FEM
Operator’s Manual and VSCC supplemental manual, and in
accordance with the requirements and sample collection filters specified
in 40 CFR Part 50, Appendix L.
[Federal
Register:
Vol 69, page 47924, 08/06/04]
Thermo Electron Model RAAS2.5-300 FEM PM2.5
Sequential Ambient Air Sampler
Manual Equivalent Method:
EQPM-0804-155
“Thermo Electron Corporation Model RAAS2.5-300 FEM” PM2.5
Sequential Ambient Air Sampler, configured with a BGI VSCC™ Very
Sharp Cut Cyclone particle size separator and operated with software
version 06.0B.00 configured for “Multi 2.5” operation, for 24-hour
continuous sample periods at a flow rate of 16.67 liters/minute, in
accordance with the Model RAAS2.5-300 FEM Operator’s Manual and
VSCC supplemental manual, and in accordance with the requirements and
sample collection filters specified in 40 CFR Part 50, Appendix
L [Federal
Register:
Vol 69, page 47924, 08/06/04]
Thermo Environmental Instruments, Incorporated Model 605 “CAPS” Sampler
Manual Reference Method:
RFPS-1098-123
“Thermo Environmental Instruments, Incorporated Model 605 “CAPS”
Computer Assisted Particle Sampler,” configured as a PM2.5
reference method and operated with software version 1.02A, for 24-hour
continuous sample periods, in accordance with the Model 605
Instruction Manual and with the requirements and sample collection
filters specified in 40 CFR Part 50, Appendix L.
[Federal
Register:
Vol 63, page 58036, 10/29/98]
PM2.5
URG-MASS100 Single PM 2.5 FRM Sampler
Manual Reference Method:
RFPS-0400-135
“URG-MASS100 Single PM 2.5 FRM Sampler,” operated with software
(firmware) version 4B or 5.0.1, configured for “Single 2.5"
operation, for 24-hour continuous sample periods at a flow rate of 16.67
liters/minute, and in accordance with the URG-MASS100
Operator’s Manual and with the requirements and sample collection
filters specified in 40 CFR Part 50, Appendix L.
[Federal
Register:
Vol 65, page 26603, 05/08/00]
URG-MASS300 Sequential PM 2.5 FRM Sampler
Manual Reference Method:
RFPS-0400-136
“URG-MASS300 Sequential PM 2.5 FRM Sampler,” operated with software
(firmware) version 4B or 5.0.1, configured for “Multi 2.5"
operation, for 24-hour continuous sample periods at a flow rate of 16.67
liters/minute, and in accordance with the URG-MASS300
Operator’s Manual and with the requirements and sample collection
filters specified in 40 CFR Part 50, Appendix L.
[Federal
Register:
Vol 65, page 26603, 05/08/00]
SULFUR DIOXIDE
Reference Method for SO2
(Pararosaniline Method)
Manual Reference Method:
40 CFR Part 50, Appendix A
Reference Method for the Determination of Sulfur Dioxide in the
Atmosphere (Pararosaniline Method)
[Federal
Register:
Vol 47, page 54899, 12/06/82 and Vol 48, 17355, 04/22/83]
Pararosaniline Method for SO2
- Technicon I
Manual Equivalent Method:
EQS-0775-001
"Pararosaniline Method for the Determination of Sulfur Dioxide in the
Atmosphere-Technicon I Automated Analysis System"
[Federal
Register:
Vol 40, page 34024, 08/13/75]
Pararosaniline Method for SO2
- Technicon II
Manual Equivalent Method:
EQS-0775-002
"Pararosaniline Method for the Determination of Sulfur Dioxide in the
Atmosphere-Technicon II Automated Analysis System"
[Federal
Register:
Vol 40, page 34024, 08/13/75]
Advanced Pollution Instrumentation, Inc. Model 100 SO2
Analyzer
Automated Equivalent Method:
EQSA-0990-077
"Advanced Pollution Instrumentation, Inc. Model 100 Fluorescent SO2
Analyzer," operated on the 0-0.1 ppm1,
the 0-0.2 ppm1,
the 0-0.5
ppm, or the 0-1.0 ppm range with a 5-micron TFE filter element installed
in the rear-panel filter assembly, either a user- or
vendor-supplied vacuum pump capable of providing 20 inches of mercury
vacuum at 2.5 L/min, with or without any of the following
options: Internal Zero/Span; Pump Pack; Rack Mount With Slides; RS-232
Interface; Status Output; TFE Zero/Span Valves; Zero Air
Scrubber; Dual Range.2
[Federal
Register:
Vol. 55, page 38149, 09/17/90]
ASARCO Model 500 SO2
Monitor
Automated Equivalent Method:
EQSA-0877-024
"ASARCO Model 500 Sulfur Dioxide Monitor," operated on a 0-0.5 ppm
range; or "ASARCO Model 600 Sulfur Dioxide Monitor,"
operated on a 0-1.0 ppm range. (Both models are identical except for the
range.) NOTE: This method is not now commercially
available. [Federal
Register:
Vol 42, page 44264, 09/02/77 and Vol 44, page 67522, 11/26/79]
Beckman Model 953 Fluorescent Ambient SO2
Analyzer
Automated Equivalent Method:
EQSA-0678-029
"Beckman Model 953 Fluorescent Ambient SO2
Analyzer," operated on a range of either 0-0.5 or 0-1.0 ppm, with a time
constant setting
of 2, 2.5, or 3 minutes, a 5 to 10 micron membrane filter element
installed in the rear-panel filter assembly, with or without any of the
following options: Remote Operation Kit, Catalog No. 641984; Digital
Panel Meter, Catalog No. 641710; Rack Mount Kit, Catalog No.
641709; Panel Mount Kit, Catalog No. 641708. [Federal
Register:
Vol 43, page 35995, 08/14/78]
Bendix Model 8303 Sulfur Analyzer
Automated Equivalent Method:
EQSA-1078-030
"Bendix Model 8303 Sulfur Analyzer," operated on a range of either 0-0.5
or 0-1.0 ppm, with a Teflon filter installed on the sample
inlet of the H2S
scrubber assembly. [Federal
Register:
Vol 43, page 50733, 10/31/78]
Columbia Scientific Industries Model 5700 SO2
Analyzer
Automated Equivalent Method:
EQSA-0494-095
"Columbia Scientific Industries Model 5700 Sulfur Dioxide Analyzer",
operated with software version 1.0 on any full scale range
between 0-250 ppb1
and 0-1000 ppb, at any integration time setting from 20 to 99 seconds,
at any temperature in the range of 15°C to
30°C, at any AC line voltage in the range of 105 to 130 volts, and with
or without any of the following options:
964-0121 Alarm Relay Contacts 964-0125 Dual Current Outputs 964-0131
Rack Mount
964-0122 Input Solenoids 964-0126 Printer 964-0012 Single Headed Pump
964-0124 Dual Analog Voltage Outputs [Federal
Register:
Vol 59, page 18818, 04/20/94]
SO2
Dasibi Model 4108 U.V. Fluorescence SO2
Analyzer
Automated Equivalent Method:
EQSA-1086-061
"Dasibi Model 4108 U.V. Fluorescence SO2
Analyzer," operated with a range of 0-100 ppb1,
0-200 ppb1,
0-500 ppb, or 0-1000 ppb, with
a Teflon-coated particulate filter and continuous hydrocarbon removal
system, with or without any of the following options: Rack
Mounting Brackets And Slides; RS-232-C Interface; Temperature
Correction. [Federal
Register:
Vol 51, page 32244, 09/10/86]
DKK-TOA Corp. Model GFS-32 U.V. Fluorescent SO2
Analyzer
Automated Equivalent Method:
EQSA-0701-115
"DKK-TOA Corporation Model GFS-32 Ambient Air SO2
Ultraviolet Fluorescent Analyzer, operated within the 0.000 to 0.500 ppm
range in the temperature range of 20°C to 30°C. [Federal
Register:
Vol 62, page 44007, 08/18/97]
DKK-TOA Corp. Model GFS-112E U. V. Fluorescent SO2
Analyzer
Automated Equivalent Method:
EQSA-0100-133
“DKK-TOA Corporation Models GFS-112E and GFS-112E-1 U.V. Fluorescence SO2
Analyzer,” operated at any temperature ranging
from 15o
C to 35o
C, on any of the following measurement ranges: 0-0.05 ppm1,
0-0.100 ppm1,
0-0.200 ppm1,
0-0.5 ppm, or 0-1.000 ppm,
and with or without the optional Internal zero air supply and permeation
tube oven.2
[Federal
Register:
Vol 65, page 2610, 01/18/00]
Environnement S.A. Model AF21M SO2
Analyzer
Automated Equivalent Method:
EQSA-0292-084
"Environnement S.A Model AF21M Sulfur Dioxide Analyzer," operated on a
range of 0-0.5 ppm with a response time coefficient setting
of 01, a Teflon filter installed in the rear-panel filter assembly, and
with or without any of the following options: Rack Mount/Slides;
RS-232-C Interface. [Federal
Register:
Vol 57, page 5444, 02/14/92]
Environnement S.A. Model AF22M SO2
Analyzer
Automated Equivalent Method:
EQSA-0802-149
“Environnement S.A Model AF22M UV Fluorescence Sulfur Dioxide Analyzer,”
operated with a full scale range of 0 - 500 ppb, at any
temperature in the range of 10 °C to 35 °C, with a 5-micron PTFE sample
particulate filter, with a response time setting of 11 (Automatic
response time), with the automatic “ZERO-REF” cycle ON and set for
activation every 24 hours, and with or without either of the
following options: Permeation oven, Rack mount slides.2
[Federal
Register:
Vol 67, page 57811, 09/12/02]
Environnement S.A. SANOA Multigas Longpath Monitoring System
Automated Equivalent Method:
EQSA-0400-138
“Environnement S.A. Model SANOA Multigas Longpath Air Quality Monitoring
System,” consisting of a receiver, one or more
projectors, interface unit, a user-provided control unit computer
running the SANOA VisionAIR software, and associated incidental
equipment; configured for measuring SO2,
with the temperature control and internal calibration cell options
installed, operated with a
measurement range of 0 to 0.5 ppm, over an installed monitoring path
length of between 27 and 500 meters, within an ambient air
temperature range of -30 to +45°C, with a measurement (integrating) time
of 180 seconds, and with or without external temperature and
barometric pressure sensors or any of the following options: external
(meteo) input connection, series 1M bus connection, OGR type
projector, analog outputs. [Federal
Register:
Vol 65, page 26603, 05/08/00]
Horiba Models APSA-360, APSA-360-CE, or APSA-360A-CE SO2
Monitors
Automated Equivalent Method:
EQSA-0197-114
"Horiba Instruments, Inc. Model APSA-360 and Model APSA-360-CE Ambient
Sulfur Dioxide Monitor,"operated with a full scale range
of 0 - 0.50 ppm, at any temperature in the range of 5 °C to 40 °C, with
a Line Setting of "MEASURE", an Analog Output Setting of
"MOMENTARY VALUE", and with or without any of the following options:2
1) Rack Mounting Plate and Side Rails, 2) RS-232 Communications Port,
and 3) Internal zero gas and span gas generator.
"Horiba Instruments, Inc. Model APSA-360A-CE Ambient Sulfur Dioxide
Monitor,” operated with one of the following measurement
ranges: 0-0.05 ppm, 0-0.1 ppm, 0-0.2 ppm, 0-0.5 ppm or 0-1.0 ppm; with
selectable time constants from 10 to 300 seconds; at any
temperature in the range of 5 °C to 40 °C; and with or without the
optional internal zero gas and span gas generator.
[Federal
Register:
Vol 62, page 6968, 02/14/97; Vol 63, page 31992, 06/11/98]
Lear Siegler Model AM2020 SO2
Monitor
Automated Equivalent Method:
EQSA-0486-049
"Lear Siegler Model AM2020 Ambient SO2
Monitor," operated on a range of either 0-0.5 or 0-1.0 ppm, at a
wavelength of 299.5 nm,
with a 5 minute integration period, over any 10°C temperature range
between 20°C and 45°C, with or without the automatic zero and
span correction feature. [Federal
Register:
Vol 45, page 79574, 12/01/80 and Vol 46, page 9997, 01/30/81]
Lear Siegler Model SM1000 SO2
Monitor
Automated Equivalent Method:
EQSA-1275-005
"Lear Siegler Model SM1000 SO2
Ambient Monitor," operated on the 0-0.5 ppm range, at a wavelength of
299.5 nm, with the "slow"
(300 second) response time, with or without any of the following
options: SM-1 Internal Zero/Span; SM-2 Span Timer Card; SM-3 0-0.1
Volt Output; SM-4 0-5 Volt Output; SM-5 Alternate Sample Pump; SM-6
Outdoor Enclosure. [Federal
Register:
Vol 41, page 3893, 01/27/76; Vol 41, page 32946, 08/06/76; Vol 42, page
13044, 03/08/77; Vol 45, page 1147, 01/04/80]
SO2
Meloy Model SA185-2A SO2
Analyzer
Automated Equivalent Method:
EQSA-1275-006
"Meloy Model SA185-2A Sulfur Dioxide Analyzer," operated on the 0-0.5
ppm range, with or without any of the following options:
S-1 Linearized Output S-2 Modified Recorder Output S-18 Rack Mount
Conversion
S-24 Dual Range Linearized Output S-5 Teflon-Coated Block S-18A Rack
Mount Conversion
S-33 Remote Range Control And Status S-6A Reignite Timer Circuit
S-21Front Panel Digital (Signals)
S-7 Press To Read Volt Meter S-34 Remote Control S-11A Manual Zero And
Span
S-22 Remote Zero/Span Control And Status (Timer) S-35 Front Panel
Digital Meter With BCD Output
S-11B Automatic Zero And Span S-13 Status Lights S-22A Remote Zero/Span
Control
S-36 Dual Range Log-Linear Output S-14 Output Booster Amplifier S-23
Automatic Zero Adjust
S-38 Sampling Mode Status S-14B Line Transmitter Board S-23A
Automatic/Manual Zero Adjust
or operated on the 0-1.0 ppm range with either option S-36 or options
S-1 and S-24, with or without any of the other options.
[Federal
Register:
Vol 41, page 3893, 01/27/76 and Vol 43, page 38088, 08/25/78]
Meloy Model SA285E SO2
Analyzer
Automated Equivalent Method:
EQSA-1078-032
"Meloy Model SA285E Sulfur Dioxide Analyzer," operated on the following
ranges and time constant switch positions:
Range, ppb: 0-501
0-1001
0-500 0-1000
Time Constant Setting: 1 or 10 1 or 10 off, 1 or 10 off, 1 or 10
The analyzer may be operated at temperatures between 10°C and 40°C and
at line voltages between 105 and 130 volts, with or without
any of the following options:
S-5 Teflon Coated Block S-22B Remote Zero/Span Control S-30 Auto
Reignite
S-14B Line Transmitter Board And Status (Pulse) S-32 Remote Range
Control And Status
S-18 Rack Mount Conversion S-23 Auto Zero Adjust S-35 Front Panel
Digital Meter With
S-18A Rack Mount Conversion S-23A Auto/Manual Zero Adjust BCD Output
S-21 Front Panel Digital Meter S-25 Press To Read S-37 Temperature
Status Lights
S-22 Remote Zero/Span Control S-26 Manual Zero And Span S-38 Sampling
Mode Status
And Status (Timer) S-27 Auto Manual Zero/Span
S-22A Remote Zero/Span Control S-28 Auto Range And Status [Federal
Register:
Vol 43, page 50733, 10/31/78]
Meloy Model SA 700 Fluorescence Sulfur Dioxide Analyzer
Automated Equivalent Method:
EQSA-0580-046
"Meloy Model SA 700 Fluorescence Sulfur Dioxide Analyzer," operated on
the 0-250 ppb1,
the 0-500 ppb, or the 0-1000 ppb range with
a time constant switch position of either 2 or 3. The analyzer may be
operated at temperatures between 20°C and 30°C and at line
voltages between 105 and 130 volts, with or without any of the following
options: FS-1 Current Output; FS-2 Rack Mount Conversion;
FS-2A Rack Mount Conversion; FS-2B Rack Mount Conversion; FS-3 Front
Panel Mounted Digital Meter; FS-5 Auto/Manual
Zero/Span With Status; FS-6 Remote/Manual Zero/Span With Status; FS-7
Auto Zero Adjust.
[Federal Register:
Vol 45, page 31488, 05/13/80]
Monitor Labs Model 8450 Sulfur Monitor
Automated Equivalent Method:
EQSA-0876-013
"Monitor Labs Model 8450 Sulfur Monitor", operated on a range of either
0-0.5 or 0-1.0 ppm, with a 5 second time constant, a model
8740 hydrogen sulfide scrubber in the sample line, with or without any
of the following options: BP Bipolar Signal Processor; IZS
Internal Zero/Span Module; V Zero/Span Valves; CLO Current Loop Output;
TF TFE Sample Particulate Filter; VT Zero/Span Valves
And Timer; DO Status Remote Interface. [Federal
Register:
Vol 41, page 36245, 08/27/76 and Vol 44, page 33476, 06/11/79]
Monitor Labs/Lear Siegler Model 8850 SO2
Analyzer
Automated Equivalent Method:
EQSA-0779-039
"Monitor Labs or Lear Siegler Model 8850 Fluorescent SO2
Analyzer," operated on a range of either 0-0.5 or 0-1.0 ppm, with an
internal
time constant setting of 55 seconds, a TFE sample filter installed on
the sample inlet line, with or without any of the following options:
03A Rack; 03B Slides; 05A Valves Zero/Span; 06A IZS Internal Zero/Span
Source; 06B,C,D NIST-Traceable Permeation Tubes; 08A
Pump; 09A Rack Mount For Option 08A; 010 Status Output W/Connector; 013
Recorder Output Options; 014 DAS Output Options;
017 Low Flow Option; 018 Kicker. [Federal
Register:
Vol 44, page 44616, 07/30/79]
Monitor Labs/Lear Siegler Model 8850S SO2
Analyzer
Automated Equivalent Method:
EQSA-0390-075
"Monitor Labs or Lear Siegler Model 8850S SO2
Analyzer," operated on a range of either 0-0.5 or 0-1.0 ppm.
[Federal
Register:
Vol 55, page 5264, 02/14/90]
Opsis Model AR 500 and System 300 Open Path Ambient Air
Automated Equivalent Method:
EQSA-0495-101
Monitoring Systems for SO2
"Opsis Model AR 500 System" or "System 300" Open Path (long path)
Ambient Air Monitoring Systems, configured for measuring SO2,
with one detector and movable grating, operated with a measurement range
of 0 to 0.5 ppm or 0 to 1.0 ppm, an installed monitoring path
length between 20 and 500 meters (or 20 and 1000 meters with the ER 150
option, AR 500 System only), xenon lamp type B (150 watt),
fiber optic cable length between 3 to 20 meters; operating within an
ambient air temperature range of -50 to +50°C, an analyzer
temperature range of 20 to 30°C, a measurement (integrating) time
setting between 30 and 120 seconds (0 min:30 sec. to 2 min:00 sec.),
SO2
and with a complete cycle time of not more than 200 seconds (3 min, 20
sec.). Under this method designation, the Model AR 500 System
or System 300 consists of: AR 500 opto-analyser; emitter EM 110 and
receiver RE 110 (together identified as ER 110); optic fibre cable
OF60-S; power supply PS 150; OPSIS operational software, version 7.0 or
7.1; and initial on-site installation, setup, and limited operator
training.2
Optional components that can be used with the Model AR 500 only,
in addition to or as alternative to corresponding components
listed above: • AR 503 opto-analyzer configured as Model AR 500 (only
the center detector active, sequential monitoring)
• Emitter/receiver ER 150 (for monitoring path lengths up to 1
kilometer) • Transceiver ER 130 and Retroreflector RE 090 with 7
prisms (max. monitoring path length 150 meters) or 12 prisms (max.
monitoring path length 250 meters) • Receiver RE 130
• Xenon lamp type A (higher short-wavelength UV output) • Optic fibre
cable OF60-R (low-loss for short wavelengths)
• Multiplexers MX 004 and MX 024 • Dataloggers DL 010 and DL 016 •
Analogue and digital input/output cards AO 008, AI 016,
and DI 032 • Analogue and digital isolation cards IA 008, ID 008, OA
008, and OD 008 • Window heaters HF 110 and HF 150
• Mirror heaters HM 110 and HM 150 • Auto calibration unit CU 007 •
Software packages IO 80 (for the analogue and digital
input/output adapters), DL10 and DL16 (for data loggers), COMVISION,
and STAT 500;
Recommended calibration and accuracy audit components (or equivalent)
for either Model AR 500 or System 300:
• Wavelength calibration lamp CA 004 • Calibration bench CB 100 •
Receiver unit RE 060 (two required) • Calibration unit CA
150, with same type lamp as used in the monitoring path emitter • Power
supply PS 150 for calibration unit CA 150 • Calibration
cells CC 001-X, where X represents various cell lengths from 1 to 900 mm
• Special calibration cells CC 110 or CC 150 (for
mounting directly on receiver) • Light meter LM 010. [Federal
Register:
Vol. 60, page 21518, 05/02/1995]
Philips PW9755 SO2
Analyzer
Automated Equivalent Method:
EQSA-0676-010
"Philips PW9755 SO2
Analyzer,"consisting of the following components: PW9755/02 SO2
Monitor with PW9741/00 SO2
Source,
PW9721/00 Filter Set SO2,
PW9711/00 Electrolyte SO2,
PW9750/00 Supply Cabinet, PW9750/10 Supply Unit/Coulometric, either
PW9731/00 Sampler or PW9731/20 Dust Filter (or vendor-approved alternate
particulate filter); operated with a 0-0.5 ppm range and
with a reference voltage setting of 760 millivolts; with or without any
of the following options: PW9750/30 Frame For MTT; PW9752/00
Air Sampler Manifold; PW9753/00 Mounting Rack For Accessories; PW9750/41
Control Clock 60 Hz; PW9754/00 Air Distributor.
[Federal
Register:
Vol 41, page 26252, 6/25/76; Vol 41, page 46019, 10/19/76; Vol 42, page
28571, 6/03/77]
Philips PW9700 SO2
Analyzer
Automated Equivalent Method:
EQSA-0876-011
"Philips PW9700 SO2
Analyzer," consisting of the following components: PW9710/00 Chemical
Unit with PW9711/00, Electrolyte SO2,
PW9721/00 Filter Set SO2,
PW9740/00 SO2
Source; PW9720/00 Electrical Unit; PW9730/00 Sampler Unit (or
vendor-approved alternate
particulate filter); operated with a 0-0.5 ppm range and with a
reference voltage of 760 millivolts.
[Federal
Register:
Vol 41, page 34105, 08/12/76]
Teledyne - Advanced Pollution Instrumentation, Inc. Models 100A, 100AS,
100E; Teledyne Analytical Instruments Model 6400A;
or Teledyne Monitor Labs sensor-e™ Model TML-50 SO2
Analyzers
Automated Equivalent Method:
EQSA-0495-100
"Teledyne - Advanced Pollution Instrumentation, Inc. Models 100A, 100AS,
or 100E; Teledyne Analytical Instruments Model 6400A;
or Teledyne Monitor Labs, Inc. sensor-e™ Model TML-50 UV Fluorescent
Sulfur Dioxide Analyzer;" operated on any full scale range
between 0-50 ppb1
and 0-1000 ppb, at any temperature in the range of 5 to 40 degrees C,
with a TFE filter element installed in the filter
assembly, with either the vendor-supplied internal pump or a user- or
vender-supplied external vacuum pump capable of maintaining
an absolute pressure of 35 cm (14 inches) of mercury (or less) at 1.0
standard liter per minute flow rate, with the following software
settings: Dynamic zero: OFF or ON; Dynamic span: OFF; AutoCal: ON or
OFF; Dual range: ON or OFF; Autorange: ON or OFF;
Temp/pressure compensation: ON; dilution factor: OFF or 1.0; and with or
without any of the following options (if available for the
various models):2
Rack mount with or without chassis slides; Fluorocarbon zero/span
valves; Internal zero/span (IZS); 4-20 mA, isolated
analog outputs; External pump; Status outputs; Control inputs; Rack
mount for external pump with tray; RS-232 output; Ethernet output;
Zero air scrubber; Combustion Filter; SO2
Permeation tube, certified or uncertified, 0.4 ppm @ 0.7 L/min; SO2
Permeation tube, certified
or uncertified, 0.8 ppm @ 0.7 L/min. [Federal
Register:
Vol. 60, page 17061, 04/04/95]
Teledyne Monitor Labs/Ecotech Model ML9850/EC9850;
Automated Equivalent Method:
EQSA-0193-092
Teledyne Monitor Labs/Ecotech/Model ML9850B/EC9850B, or Wedding &
Associates Model 1040 SO2
Analyzers
"Teledyne Monitor Labs or Ecotech Models ML9850/EC9850 or
ML9850B/EC9850B, or Wedding & Associates, Inc. Model 1040
Sulfur Dioxide Analyzers," operated on any full scale range between
0-0.050 ppm1
and 0-1.0 ppm, at any temperature in the range of
15°C to 35°C, with the service switch on the secondary panel set to the
In
position; with the following menu choices selected: Range:
0.05 ppm to 1.0 ppm;
Over-ranging:
Enabled
or
Disabled;
Background:
Not Disabled;
Calibration:
Manual
or
Timed;
Diagnostic Mode:
Operate;
Filter Type:
Kalman;
Pres/Temp/Flow Comp:
On;
Span Comp:
Disabled;
and as follows:
Model ML9850/EC9850
- with a
five-micron Teflon® filter element installed internally, with the 50-pin
I/O board installed on the rear panel configured at any of the
following output range settings: Voltage, 0.1 V, 1 V, 5 V, 10 V;
Current, 0-20 mA, 2-20 mA, 4-20 mA; and with or without any of the
following options: Valve Assembly for External Zero/Span (EZS); Rack
Mount Assembly; Internal Floppy Disk Drive.
Models
ML9850B/EC9850B and 1040
- with either a vendor-supplied or equivalent user supplied five-micron
Teflon® filter, zero air scrubber,
SO2...O3
and exhaust pump, and with or without any of the following options:
Valve Assembly for External Zero/Span (EZS); Rack Mount
Assembly; 50-pin I/O board; Exhaust Scrubber; Internal Zero/Span
Assembly (IZS); hinged, fold-down front panel.
[Federal
Register:
Vol 58, page 6964, 02/03/93]
Thermo Electron Model 43 SO2
Analyzer
Automated Equivalent Method:
EQSA-0276-009
"Thermo Electron Model 43 Pulsed Fluorescent SO2
Analyzer," equipped with an aromatic hydrocarbon cutter and operated on
a range
of either 0-0.5 or 0-1.0 ppm, with or without any of the following
options: 001 Rack Mounting For Standard 19 Inch Relay Rack; 002
Automatic Actuation Of Zero And Span Solenoid Valves; 003 Type S Flash
Lamp Power Supply; 004 Low Flow.
[Federal
Register:
Vol 41, page 8531, 02/27/76; Vol 41, page 15363, 04/12/76; Vol 42, page
20490, 04/20/77
Vol 44, page 21861, 04/12/79; Vol 45, page 2700, 01/14/80; Vol 45, page
32419, 05/16/80]
Thermo Environmental Instruments, Inc. Models 43A, 43B, 43C Analyzers
Automated Equivalent Method:
EQSA-0486-060
"Thermo Electron or Thermo Environmental Instruments, Inc. Model 43A or
43B Pulsed Fluorescence SO2
Analyzer," operated on the
0-0.1 ppm1,
the 0-0.2 ppm1,
the 0-0.5 ppm, or the 0-1.0 ppm range, with either a high or a low time
constant setting (Model 43A) and
with or without any of the following options:2
001 Teflon Particulate Filter 004 High Flow Rate (1 LPM) 007 Remote
Activation Of Zero/Span Valves
002 19" Rack Mounting Configuration 005 Current Output 008 RS-232
Interface (Model 43B)
003 Internal Zero/Span Valves 006 Internal Permeation 009
Pressure/Temperature Compensation
Span Source (Model 43B)
"Thermo Environmental Instruments, Inc. Model 43C Pulsed Fluorescence SO2
Analyzer," operated on any measurement range between
0-50 ppb1
and 0-1000 ppb, with any time average setting from 10 to 300 seconds,
with temperature and/or pressure compensation on or
off, operated at temperatures between 20
oC
and 30
oC,
with or without any of the following options:2
43C-001 Teflon particulate filter 43C-005 4-20 mA current output
43C-002 Rack mounts 43C-006 Internal permeation span source
43C-003 Internal zero/span and sample solenoid valves 43C-007 Remote
activation of zero/span and sample valves
43C-004 High flow rate (0.5-1.0 LPM) 43C-008 RS-232/485 interface
[Federal
Register:
Vol 51, page 12390, 04/10/86]
OZONE
Beckman Model 950A Ozone Analyzer
Automated Reference Method:
RFOA-0577-020
"Beckman Model 950A Ozone Analyzer," operated on a range of 0-0.5 ppm
and with the "SLOW" (60 second) response time, with or
without any of the following options: Internal Ozone Generator; Computer
Adaptor Kit; Pure Ethylene Accessory.
[Federal
Register:
Vol 42, page 28571, 06/03/77]
Bendix or Combustion Engineering Model 8002 Ozone Analyzer
Automated Reference Method:
RFOA-0176-007
"Bendix or Combustion Engineering Model 8002 Ozone Analyzer", operated
on the 0-0.5 ppm range, with a 40 second time constant,
with or without any of the following options: Rack Mounting With Chassis
Slides; Rack Mounting Without Chassis Slides; Zero And
Span Timer; Ethylene/CO2
Blend Reactant Gas. [Federal
Register:
Vol 41, page 5145, 02/04/76 and Vol 45, page 18474, 03/21/80]
Columbia Scientific Industries Model 2000 Ozone Meter
Automated Reference Method:
RFOA-0279-036
"Columbia Scientific Industries Model 2000 Ozone Meter," when operated
on the 0-0.5 ppm range with either AC or battery power:
The BCA 952 battery charger/AC adapter M952-0002 (115V) or M952-0003
(230V) is required for AC operation; an internal battery
M952-0006 or 12 volt external battery is required for portable non-AC
powered operation.
[Federal
Register:
Vol 44, page 10429, 02/20/79]
Dasibi Models 1003-AH, 1003-PC, or 1003-RS Ozone Analyzers
Automated Equivalent Method:
EQOA-0577-019
"Dasibi Model 1003-AH, 1003-PC, or 1003-RS Ozone Analyzer," operated on
a range of either 0-0.5 or 0-1.0 ppm, with or without any
of the following options: Adjustable Alarm; Aluminum Coated Absorption
Tubes, Integrated Output; Vycor-Jacketed U.V. Source
Lamp; BCD Digital Output; Rack Mounting Ears And Slides; 0-10 mV, 0-100
mV, 0-1 V, Or 0-10 V; Glass (Pyrex) Absorption Tubes;
Teflon-based Solenoid Valve; Analog Output. [Federal
Register:
Vol 42, page 28571, 06/03/77]
O3
Dasibi Models 1008-AH, 1008-PC, or 1008-RS Ozone Analyzers
Automated Equivalent Method:
EQOA-0383-056
"Dasibi Model 1008-AH, 1008-PC, or 1008-RS Ozone Analyzer," operated on
a range of either 0-0.5 or 0-1.0 ppm, with or without any
of the following options: Aluminum Coated Absorption Tubes; BCD Digital
Output; RS232 Interface; Glass (Pyrex) Absorption Tubes;
Vycor-Jacketed U.V. Source Lamp; Ozone Generator; Teflon-based Solenoid
Valve; Photometer Flow Restrictor (2 LPM); 4-20 mA,
Isolated, Or Dual Analog Outputs; Rack Mounting Brackets Or Slides; 20
Second Update Software.
[Federal
Register:
Vol 48, page 10126, 03/10/83]
DKK-TOA Corp. Model GUX-113E Ozone Analyzer
Automated Equivalent Method:
EQOA-0200-134
“DKK-TOA Corporation Models GUX-113E and GUX-113E-1 Ozone Analyzer,”
operated at any temperature in the range of 15o
C to
35o
C, on any of the following measurement ranges: 0-0.100 ppm1,
0-0.200 ppm1,
0-0.5 ppm, or 0-1.000 ppm, and with or without the
optional Internal Ozone Generator.2
[Federal
Register:
Vol 65, page 11308, 03/02/00]
Environics Series 300 Ozone Analyzer
Automated Equivalent Method:
EQOA-0990-078
"Environics Series 300 Computerized Ozone Analyzer," operated on the
0-0.5 ppm range, with the following parameters entered into
the analyzer's computer system: Absorption Coefficient =
308 ± 4;
Flush Time =
3;
Integration Factor =
1;
Offset Adjustment =
0.025
ppm;
Ozone Average Time =
4;
Signal Average =
0;
Temp/Press Correction =
On;
and with or without the RS-232 Serial Data Interface.
[Federal
Register:
Vol 55, page 38386, 09/18/90]
Environnement S.A. Model O341M
UV Ozone Analyzer
Automated Equivalent Method:
EQOA-0895-105
"Environnement S.A. Model O341M
UV Photometric Ozone Analyzer," operated on a full scale range of 0 -
500 ppb, at any temperature
in the range of 15 °C to 35 °C, with the response time set to 50
seconds, and with or without any of the following options:2
Internal
Ozone Generator; Span External Control; RS232-422 Serial Interface;
Internal Printer.
[Federal
Register:
Vol. 60, page 39382, 08/02/95]
Environnement S.A. Model O342M
UV Ozone Analyzer
Automated Equivalent Method:
EQOA-0206-148
“Environnement S.A Model O342M
UV Photometric Ozone Analyzer,” operated with a full scale range of 0 -
500 ppb, at any
temperature in the range of 10 °C to 35 °C, with a 5-micron PTFE sample
particulate filter, with response time setting of 11 (Automatic
response time), and with or without any of the following options:2
c) Internal ozone generator, d) Span external control (zero/span
solenoid valve). [Federal
Register:
Vol. 67, page 42557, 06/24/02]
Environnement S.A. SANOA Multigas Longpath Monitoring System
Automated Equivalent Method:
EQOA-0400-137
“Environnement S.A. Model SANOA Multigas Longpath Air Quality Monitoring
System, consisting of a receiver, one or more
projectors, interface unit, a user-provided control unit computer
running the SANOA VisionAIR software, and associated incidental
equipment; configured for measuring O3,
with the temperature control and internal calibration cell options
installed, operated with a
measurement range of 0 to 0.5 ppm, over an installed monitoring path
length of between 27 and 500 meters, within an ambient air
temperature range of -30 to +45°C, with a measurement (integrating) time
of 180 seconds, and with or without external temperature and
barometric pressure sensors or any of the following options: external
(meteo) input connection, series 1M bus connection, OGR type
projector, analog outputs. A high-concentration ozone generator, part #
80-231-03, or the SONIMIX 7121B calibration system is
recommended for calibration or accuracy auditing [Federal
Register:
Vol 65, page 26603, 05/08/00]
Horiba Instruments Models APOA-360 and APOA-360-CE Ozone Monitor
Automated Equivalent Method:
EQOA-0196-112
"Horiba Instruments, Inc. Model APOA-360 or APOA-360-CE Ambient Ozone
Monitor," operated with a full scale range of 0 - 0.50
ppm, at any temperature in the range of 10°C to 40°C, with a Line
Setting of "MEASURE", and an Analog Output of "MOMENTARY
VALUE", and with or without any of the following options:2
1) Rack Mounting Plate and Side Rails 2) RS-232 Communications Port,
and 3) Optional Internal Zero/Span Check [Federal
Register:
Vol. 61, page 11404, 03/20/96]
McMillan (MEC) Models 1100-1, 1100-2, and 1100-3 Ozone Meters
"MEC Model 1100-1 Ozone Meter,"
Automated Reference Method:
RFOA-1076-014
"MEC Model 1100-2 Ozone Meter,"
Automated Reference Method:
RFOA-1076-015
"MEC Model 1100-3 Ozone Meter,"
Automated
Reference Method:
RFOA-1076-016
operated on a 0-0.5 ppm range, with or without any of the following
options: 0011 Rack Mounting Ears; 0026 Alarm Set Feature; 0012
Instrument Bail; 0033 Local-Remote Sample; Zero, Span Kit Blend Feature;
0016 Chassis Slide Kit; 0040 Ethylene/CO2.
[Federal
Register:
Vol 41, page 46647, 10/22/76 and Vol 42, page 30235, 06/13/77]
Meloy Model OA325-2R Ozone Analyzer
Automated Reference Method:
RFOA-1075-003
"Meloy Model OA325-2R Ozone Analyzer," operated with a scale range of
0-0.5 ppm, with or without any of the following options:
0-4 Output Booster Amplifier; 0-18 Rack Mount Conversion; 0-18A Rack
Mount Conversion.
[Federal
Register:
Vol 40, page 54856, 11/26/75]
O3
Meloy Model OA350-2R Ozone Analyzer
Automated Reference Method:
RFOA-1075-004
"Meloy Model OA350-2R Ozone Analyzer," operated with a scale range of
0-0.5 ppm, with or without any of the following options:
0-2 Automatic Zero And Span; 0-3 Remote Control Zero And Span; 0-4
Output Booster Amplifier; 0-18 Rack Mount Conversion;
0-18A Rack Mount Conversion. [Federal
Register:
Vol 40, page 54856, 11/26/75]
Monitor Labs Model 8410E Ozone Analyzer
Automated Reference Method:
RFOA-1176-017
"Monitor Labs Model 8410E Ozone Analyzer," operated on a range of 0-0.5
ppm with a time constant setting of 5 seconds, with or
without any of the following options: DO Status Outputs; ER Ethylene
Regulator Assembly; V TFE Zero/Span Valves; TF TFE
Sample Particulate Filter; VT TFE Zero/Span Valves And Timer. [Federal
Register:
Vol 41, page 53684, 12/08/76]
Monitor Labs/Lear Siegler Model 8810 Ozone Analyzer
Automated Equivalent Method:
EQOA-0881-053
"Monitor Labs or Lear Siegler Model 8810 Photometric Ozone Analyzer,"
operated on a range of either 0-0.5 or 0-1.0 ppm, with
selectable electronic time constant settings from 20 through 150
seconds, with or without any of the following options: 05 Pressure
Compensation; 06 Averaging Option; 07 Zero/Span Valves; 08 Internal
Zero/Span (Valve And Ozone Source); 09 Status; 10
Particulate Filter; 15 through 20 DAS/REC Output. [Federal
Register:
Vol 46, page 52224, 10/26/81]
Opsis Model AR 500 and System 300 Open Path Ambient Air
Automated Equivalent Method:
EQOA-0495-103
Monitoring Systems for Ozone
"Opsis Model AR 500 System" or "System 300" Open Path (long path)
Ambient Air Monitoring Systems, configured for measuring O3,
with one detector and moveable grating, operated with a measurement
range of 0 to 0.5 ppm, an installed monitoring path length between
20 and 500 meters (or 20 and 1000 meters with the ER 150 option, AR 500
System only), xenon lamp type B (150 watt), fiber optic cable
length between 3 to 20 meters; operating within an ambient air
temperature range of -50 to +50°C, an analyzer temperature range of
20 to 30°C, a measurement (integrating) time setting between 30 and 120
seconds (0 min:30 sec. to 2 min:00 sec.), and with a complete
cycle time of not more than 200 seconds (3 min, 20 sec.). Under this
method designation, the Model AR 500 System or System 300
consists of: AR 500 opto-analyser; emitter EM 110 and receiver RE 110
(together identified as ER 110); optic fibre cable OF60-S; power
supply PS 150, OPSIS operational software, version 7.0 or 7.1; and
initial on-site installation, setup, and limited operator training.2
Optional components that can be used with the Model AR 500 only,
in addition to or as alternative to corresponding components
listed above: • AR 503 opto-analyzer configured as Model AR 500 (only
the center detector active, sequential monitoring)
• Emitter/receiver ER 150 (for monitoring path lengths up to1 kilometer)
• Transceiver ER 130 and Retroreflector RE 090 with 7
prisms (max. monitoring path length 150 meters) or 12 prisms (max.
monitoring path length 250 meters) • Receiver RE 130 • Optic
fibre cable OF60-R (low-loss for short wavelengths) • Multiplexers MX
004 and MX 024 • Dataloggers DL 010 and DL 016
• Analogue and digital input/output cards AO 008, AI 016, and DI 032 •
Analogue and digital isolation cards IA 008, ID 008,
OA 008, and OD 008 • Window heaters HF 110 and HF 150 • Mirror heaters
HM 110 and HM 150 • Auto calibration unit CU 007
• Software packages IO 80 (for the analogue and digital input/output
adapters), DL10 and DL16 (for data loggers), ComVision, and
STAT 500;
Recommended calibration and accuracy audit components (or equivalent)
for either Model AR 500 or System 300:
• Wavelength calibration lamp CA 004 • Calibration bench CB 100 •
Receiver unit RE 060 (two required) • Calibration unit CA
150, with same type lamp as used in the monitoring path emitter • Power
supply PS 150 for calibration unit CA 150 • Calibration
cells CC 001-X, where X represents various cell lengths from 1 to 900 mm
• Special calibration cells CC 110 or CC 150 (for
mounting directly on receiver) • Ozone generator OC 500 • Light meter LM
010.
Federal Register:
Vol. 60, page 21518, 05/02/1995]
PCI Ozone Corporation Model LC-12 Ozone Analyzer
Automated Equivalent Method:
EQOA-0382-055
"PCI Ozone Corporation Model LC-12 Ozone Analyzer," operated on a range
of 0-0.5 ppm.
[Federal
Register:
Vol 47, page 13572, 03/31/82]
Philips PW9771 03 Analyzer
Automated Equivalent Method:
EQOA-0777-023
"Philips PW9771 03 Analyzer," consisting of the following components:
PW9771/00 03 Monitor with PW9724/00 Disc.-Set;
PW9750/00 Supply Cabinet; PW9750/20 Supply Unit operated on a range of
0-0.5 ppm, with or without any of the following
accessories: PW9732/00 Sampler Line Heater; PW9750/30 Frame For MTT;
PW9750/41 Control Clock 60 Hz; PW9733/00 Sampler;
PW9752/00 Air Sampler Manifold. [Federal
Register:
Vol 42, page 38931, 08/01/77; Vol 42, page 57156, 11/01/77]
Teledyne - Advanced Pollution Instrumentation, Inc. Model 400E; Advanced
Pollution Instrumentation, Inc. Model 400/400A;
Teledyne Monitor Labs sensor-e™ Model TML-10 Ozone Analyzers
Automated Equivalent Method:
EQOA-0992-087
"Teledyne - Advanced Pollution Instrumentation, Inc. Model 400E;
Advanced Pollution Instrumentation, Inc. Model 400 or 400A; or
Teledyne Monitor Labs sensor-e™ Model TML-10 Ozone Analyzer;" operated
on any full scale range between 0-100 ppb1
and 0-1000
ppb, with any range mode (Single, Dual, or AutoRange), at any ambient
temperature in the range of 5°C to 40°C, and with a TFE filter.
Models 400E and TML-50:
operated with a sample flow rate of 800 ±80 cm3/min
(sea level), with the dilution factor set to 1, with
Dynamic Zero ON or OFF, with Dynamic Span OFF, with Temp/Press
compensation ON, and with or without any of the following
options: Internal or external sample pump, Sample/Cal valve option,
Internal Zero/Span (IZS), Rack mount with or without slides, 4-20
O3
mA isolated current loop output.2
Models 400/400A:
operated with the dynamic zero and span adjustment feature (some Model
400
units only) set to OFF, and with or without any of the following
options: Zero/Span Valve option, Internal Zero/Span (IZS) option, IZS
ozone generator reference feedback option, standard serial port or
Multi-drop RS-232, digital status outputs, analog outputs: 100 mV,
1 V, 5 V, 10 V, 4-20 mA current loop, optional metal wool ozone
scrubber, optional external sample pump, optional 47 mm diameter
filter, optical bench heater, rack mount with slides.
[Federal
Register:
Vol 57, page 44565, 09/28/92; Vol 63, page 31992, 06/11/98;
Federal Register:
Vol 67, page 57811, 09/12/02]
Teledyne Monitor Labs/Ecotech Models ML9810/EC9810, -9811, or -9812,
Automated Equivalent Method:
EQOA-0193-091
Teledyne Monitor Labs/Ecotech Model ML9810B/EC9810B, or Wedding &
Associates Model 1010 Ozone Analyzers
"Teledyne Monitor Labs or Ecotech Models ML9810/EC9810, ML9811/EC9811,
ML9812/EC9812, or ML9810B/EC9810B or Wedding
& Associates, Inc. Model 1010 Ozone Analyzers," operated on any full
scale range between 0-0.05 ppm1
and 0-1.0 ppm, at any
temperature in the range of 15°C to 35°C, with the service switch on the
secondary panel set to the
In
position; with the following menu
choices selected: Range:
0.05 ppm
to
1.0 ppm;
Over-ranging:
Enabled
or
Disabled;
Calibration:
Manual
or
Timed;
Diagnostic Mode:
Operate;
Filter Type:
Kalman;
Pres/Temp/Flow Comp:
On;
Span Comp:
Disabled;
and as follows:
Models ML9810/EC9810, -9811,
and -9812
- with a five-micron Teflon® filter element installed internally, with
the 50-pin I/O board installed on the rear panel configured
at any of the following output range settings: Voltage, 0.1 V, 1 V, 5 V,
10 V; Current, 0-20 mA, 2-20 mA, 4-20 mA; and with or without
any of the following options: Valve Assembly for External Zero/Span
(EZS); Rack Mount Assembly; Internal Floppy Disk Drive.
Models ML9810B/EC9810B and 1010
- with either a vendor-supplied or equivalent user-supplied five micron
Teflon® filter and
exhaust pump, and with or without any of the following options: Valve
Assembly for External Zero/Span (EZS); Rack Mount Assembly;
50-pin I/O board; Internal Zero/Span Assembly (IZS); hinged, fold-down
front panel.
[Federal
Register:
Vol 58, page 6964, 02/03/93]
Thermo Electron/Thermo Environmental Instruments Models 49, 49C
Automated Equivalent Method:
EQOA-0880-047
"Thermo Electron or Thermo Environmental Instruments, Inc. Model 49 U.V.
Photometric Ambient O3
Analyzer" operated on a
measurement range of either 0-0.5 or 0-1.0 ppm with or without any of
the following options:
49-001 Teflon Particulate Filter; 49-002 19 Inch Rack Mount; 49-100
Internal Ozone Generator for Zero, Precision, and Level
1 Span Check; 49-103 Internal Ozone Generator for Zero, Precision, and
Level 1 Span Checks With Remote Activation;
49-488 GPIB (General Purpose Interface Bus) IEEE-488
"Thermo Electron or Thermo Environmental Instruments, Inc. Model 49C
U.V. Photometric Ambient O3
Analyzer" operated on any
measurement range between 0-0.051
to 1.0 ppm with any time average setting between 10 and 300 seconds,
with the temperature and/or
pressure compensation on or off, with or without any of the following
options:
2
100 Teflon particulate filter 420 Internal Zero Air Scrubber
200 Carrying Handle 610 4-20 mA current output
210 Rack mounts 730 RS-232 Interface
340 Internal Ozonator 780 RS-485 Interface
350 Internal Ozonator with Remote I/O Activation [Federal
Register:
Vol 45, page 57168, 08/27/80]
CARBON MONOXIDE
Beckman Model 866 CO Monitoring System
Automated Reference Method:
RFCA-0876-012
"Beckman Model 866 Ambient CO Monitoring System," consisting of the
following components: Pump/Sample-Handling Module; Gas
Control Panel; Model 865-17 Analyzer Unit; Automatic Zero/Span
Standardizer; operated with a 0-50 ppm range, a 13 second electronic
response time, with or without any of the following options: Current
Output Feature; Bench Mounting Kit; Linearizer Circuit.
[Federal
Register:
Vol. 41, page 36245, 08/27/76]
Bendix/Combustion Engineering Model 8501-5CA CO Analyzer
Automated Reference Method:
RFCA-0276-008
"Bendix or Combustion Engineering Model 8501-5CA Infrared CO Analyzer",
operated on the 0-50 ppm range and with a time constant
setting between 5 and 16 seconds, with or without any of the following
options: Rack Mounting With Chassis Slides; Rack Mounting
Without Chassis Slides; External Sample Pump. [Federal
Register:
Vol. 41, page 7450, 02/18/76]
Dasibi Model 3003 CO Analyzer
Automated Reference Method:
RFCA-0381-051
"Dasibi Model 3003 Gas Filter Correlation Dasibi Environmental CO
Analyzer," operated on the 0-50 ppm range, with a sample
particulate filter installed on the sample inlet line, with or without
any of the following options:
3-001 Rack Mount 3-003 BCD Digital Output 3-007 Zero/Span Module Panel
3-002 Remote Zero And Span 3-004 4-20 Milliamp Output [Federal
Register:
Vol. 46, page 20773, 04/07/81]
CO
Dasibi Model 3008 CO Analyzer
Automated Reference Method:
RFCA-0488-067
"Dasibi Model 3008 Gas Filter Correlation CO Analyzer," operated on the
0-50 ppm range, with a time constant setting of 60 seconds,
a particulate filter installed in the analyzer sample inlet line, with
or without use of the auto zero or auto zero/span feature, and with or
without any of the following options: N-0056-A RS-232-C Interface;
S-0132-A Rack Mounting Slides; Z-0176-S Rack Mounting
Brackets. [Federal
Register:
Vol. 53, page 12073, 04/12/88]
Environnement S.A. Model CO11M CO Analyzer
Automated Reference Method:
RFCA-0995-108
"Environnement S.A. Model CO11M Ambient Carbon Monoxide Analyzer,"
operated on a full scale range of 0 - 50 ppm, at any
temperature in the range of 15 °C to 35 °C, with a 5-micron PTFE sample
particulate filter, with the following software settings:
Automatic response time ON; Minimum response time set to 40 seconds (RT
13); Automatic ZERO-REF cycle programmed every 24
hours; and with or without any of the following options:
2
RS232-422 Serial Interface; Internal Printer.
[Federal
Register:
Vol. 60, page 54684, 10/25/95]
Environnement S.A. Model CO12M CO Analyzer
Automated Reference Method:
RFCA-0206-147
“Environnement S.A Model CO12M Gas Filter Correlation Carbon Monoxide
Analyzer,” operated with a full scale range of 0 - 50 ppm,
at any temperature in the range of 10 °C to 35 °C, with a 5-micron PTFE
sample particulate filter, with response time ON, and with the
automatic “ZERO-REF” cycle either ON or OFF.2
[Federal
Register:
Vol. 67, page 42557, 06/24/02]
Horiba Models AQM-10, AQM-11, and AQM12 CO Monitoring Systems
Automated Reference Method:
RFCA-1278-033
"Horiba Models AQM-10, AQM-11, and AQM12 Ambient CO Monitoring Systems,"
operated on the 0-50 ppm range, with a response
time setting of 15.5 seconds, with or without any of the following
options: AIC-101 Automatic Indication Corrector; VIT-3 Non-Isolated
Current Output; ISO-2 And DCS-3 Isolated Current Output. [Federal
Register:
Vol. 43, page 58429, 12/14/78]
Horiba Model APMA-300E CO Monitoring System
Automated Reference Method:
RFCA-1180-048
"Horiba Model APMA-300E Ambient Carbon Monoxide Monitoring System,"
operated on the 0-20 ppm1,
the 0-50 ppm, or the 0-100
ppm range with a time constant switch setting of No. 5. The monitoring
system may be operated at temperatures between 10°C and 40°C.
(This method was originally designated as "Horiba Model APMA 300E/300SE
Ambient Carbon Monoxide Monitoring System".)
[Federal
Register:
Vol. 45, page 72774, 11/03/80]
Horiba Model APMA-360 CO Monitor
Automated Reference Method:
RFCA-0895-106
"Horiba Instruments Incorporated, Model APMA-360 Ambient Carbon Monoxide
Monitor," operated on the 0-50 ppm range, with the
Line Setting set to "MEASURE", with the Analog Output set to "MOMENTARY
VALUE", and with or without the following options:2
1) Rack Mounting Plate and Side Rails 2) RS-232 Com Port. [Federal
Register:
Vol. 60, page 39382, 08/02/95]
MASS-CO, Model 1 CO Analyzer
Automated Reference Method:
RFCA-1280-050
"MASS-CO, Model 1 Carbon Monoxide Analyzer," operated on a range of 0-50
ppm, with automatic zero and span adjustments at time
intervals not to exceed 4 hours, with or with out the 100 millivolt and
5 volt output options. The method consists of the following
components: (1) Infra-2 (Uras 2) Infrared Analyzer Model 5611-200-35,
(2) Automatic Calibrator Model 5869-111, (3) Electric Gas
Cooler Model 7865-222 or equivalent with prehumidifier, (4) Diaphragm
Pump Model 5861-214 or equivalent, (5) Membrane Filter
Model 5862-111 or equivalent, (6) Flow Meter Model SK 1171-U or
equivalent, (7) Recorder Model Mini Comp DN 1/192 or
equivalent. NOTE: This method is not now commercially available. [Federal
Register:
Vol. 45, page 81650, 12/11/80]
Monitor Labs Model 8310 CO Analyzer
Automated Reference Method:
RFCA-0979-041
"Monitor Labs Model 8310 CO Analyzer," operated on the 0-50 ppm range,
with a sample inlet filter, with or without any of the
following options: 02A Zero/Span Valves 04B Pump (50 Hz) 07A Zero/Span
Valve Power Supply
03A Floor Stand 05A CO Regulator 08A Calibration Valves
04A Pump (60 Hz) 06A CO Cylinder 9A,B,C,D Input Power Transformer
[Federal
Register:
Vol. 44, page 54545, 09/20/79 and Vol. 45, page 2700, 01/14/80]
Monitor Labs/Lear Siegler Model 8830 CO Analyzer
Automated Reference Method:
RFCA-0388-066
"Monitor Labs or Lear Siegler Model 8830 CO Analyzer," operated on the
0-50 ppm range, with a five micron Teflon filter element
installed in the rear-panel filter assembly, with or without any of the
following options: 2 - Zero/Span Valve Assembly; 3 - Rack
Assembly; 4 - Slide Assembly; 7 - 230 VAC, 50/60 Hz. [Federal
Register:
Vol. 53, page 7233, 03/07/88]
MSA/LIRA Model 202S CO Analyzer System
Automated Reference Method:
RFCA-0177-018
"LIRA Model 202S Air Quality Carbon Monoxide Analyzer System,"
consisting of a LIRA Model 202S optical bench (P/N 459839),
a regenerative dryer (P/N 464084), and rack-mounted sampling system;
operated on a 0-50 ppm range, with the slow response amplifier,
with or without any of the following options: Remote Meter; Remote Zero
And Span Controls; 0-1, 5, 20, Or 50 mA Output; 1-5, 4-20,
Or 10-50 mA Output; 0-10 Or 100 mV Output; 0-1, 5, Or 10 Volt Output. [Federal
Register:
Vol. 42, page 5748, 01/31/77]
CO...NO2
Teledyne Advanced Pollution Instrumentation, Inc. Models 300 or 300E or
Teledyne Monitor Labs sensor-e™ Model TML-30 CO Analyzer
Automated Reference Method:
RFCA-1093-093
"Teledyne Advanced Pollution Instrumentation, Inc. Models 300 or 300E,
or Teledyne Monitor Labs, Inc. sensor-e™ Model TML-30
Gas Filter Correlation Carbon Monoxide Analyzer," operated on any full
scale range between 0-10 ppm and 0-50 ppm, at any
temperature in the range of 15°C to 35°C for
Model 300
or 10°C to 40°C for
Models 300E and TML-30,
with a 5-micron TFE filter
element installed in the filter assembly, with the dynamic zero and span
adjustment set to
Off
for
Model 300,
and with or without any
of the following options2:
Option 50, Zero/Span Valves with pressurized span gas and shutoff valve;
Option 51, Zero/Span Valves with
pressurized span gas and shutoff valve and Internal Zero Air Generator;
Option 52, Zero/Span Valves; Option 53, Zero/Span Valves with
Internal Zero Air Generator; Rack Mount with slides; RS-232 serial port
with status outputs; and (for
Models 300E and TML-30)
4-20
mA isolated outputs. [Federal
Register:
Vol. 58, page 58166, 10/29/93]
Teledyne Monitor Labs/Ecotech/Model ML9830/EC9830,
Automated Reference Method:
RFCA-0992-088
Teledyne Monitor Labs/Ecotech Model ML9830B/EC9830B, or Wedding &
Associates Model 1020 CO Analyzers
"Teledyne Monitor Labs or Ecotech Models ML9830/EC9830 or
ML9830B/EC9830B or Wedding & Associates, Inc. Model 1020
Carbon Monoxide Analyzer," operated on any full scale range between
0-5.0 ppm1
and 0-100 ppm, at any temperature in the range of
15°C to 35°C, with the service switch on the secondary panel set to the
In
position, with the following menu choices selected: Range:
5.0 ppm
to
100.0
ppm; Over-ranging:
Enabled
or
Disabled;
Background:
Not Disabled;
Calibration:
Manual
or
Timed;
Diagnostic Mode:
Operate;
Filter Type:
Kalman;
Pres/Temp/Flow Comp:
On;
Span Comp:
Disabled;
and as follows:
Model ML9830/EC9830:
with a
five-micron Teflon® filter element installed internally, with the 50-pin
I/O board installed on the rear panel configured at any of the
following output range settings: Voltage, 0.1 V, 1 V, 5 V, 10 V;
Current, 0-20 mA, 2-20 mA and 4-20 mA; and with or without any of
the following options: Valve Assembly for External Zero/Span (EZS);
Valve Assembly for Internal Zero/Span (IZS); Rack Mount
Assembly; Internal Floppy Disk Drive.
Models ML9830B/EC9830B and 1020:
with either a vendor-supplied or equivalent usersupplied
five micron Teflon® filter and exhaust pump, and with or without any of
the following options: Valve Assembly for External
Zero/Span (EZS); 50-pin I/O board; Rack Mount Assembly; High Pressure
Span Valve; hinged, fold-down front panel.
[Federal
Register:
Vol. 57, page 44565, 09/28/92]
Thermo Electron/Thermo Environmental Instruments Models 48, 48C
Automated Reference Method:
RFCA-0981-054
"Thermo Electron or Thermo Environmental Instruments, Inc. Model 48 Gas
Filter Correlation Ambient CO Analyzer," operated on the
0-50 ppm range, with a time constant setting of 30 seconds, with or
without any of the following options:
48-001 Teflon Particulate Filter 48-010 Internal Zero Air Package
48-002 19 Inch Rack Mount 48-488 GPIB (General Purpose Interface Bus)
EEEE-488
48-003 Internal Zero/Span Valves with Remote Activation
"Thermo Electron or Thermo Environmental Instruments, Inc. Model 48C Gas
Filter Correlation Ambient CO Analyzer," operated on
any measurement range between 0-1 ppm1
and 0-100 ppm, with any time average setting from 10 to 300 seconds,
with temperature
and/or pressure compensation on or off, operated at temperatures between
20
oC
and 30
oC,
with or without any of the following
options:2
100 Teflon particulate filter 410 Internal Zero Air Scrubber
200 Carrying Handle 610 4-20 mA current output
210 Rack mounts 720 RS-232 Interface
320 Internal Zero/Span and Sample/Calibration Solenoid Valves 770 RS-485
Interface
330 Internal Zero/Span and Sample/ Calibration Solenoid Valves
with Remote I/O Activation [Federal
Register:
Vol. 46, page 47002, 09/23/81]
NITROGEN DIOXIDE
Sodium Arsenite Method for NO2
Manual Equivalent Method:
EQN-1277-026
"Sodium Arsenite Method for the Determination of Nitrogen Dioxide in the
Atmosphere."
[Federal
Register:
Vol. 42, page 62971, 12/14/77]
Sodium Arsenite Method for NO2
- Technicon II
Manual Equivalent Method:
EQN-1277-027
"Sodium Arsenite Method for the Determination of Nitrogen Dioxide in the
Atmosphere-Technicon II Automated Analysis System."
[Federal
Register:
Vol. 42, page 62971, 12/14/77]
TGS-ANSA Method for NO2
Manual Equivalent Method:
EQN-1277-028
"TGS-ANSA Method for the Determination of Nitrogen Dioxide in the
Atmosphere."
[Federal
Register:
Vol. 42, page 62971, 12/14/77]
NO2
Advanced Pollution Instrumentation, Inc. Model 200 NO2
Analyzer
Automated Reference Method:
RFNA-0691-082
"Advanced Pollution Instrumentation, Inc. Model 200 Nitrogen Oxides
Analyzer," operated on a range of either 0-0.5 or 0-1.0 ppm, with
a 5-micron TFE filter element installed in the rear-panel filter
assembly, with either a user- or vendor-supplied vacuum pump capable
of providing 5 inches mercury absolute pressure at 5 slpm, with either a
user- or vendor-supplied dry air source capable of providing
air at a dew point of 0°C or lower, with the following settings of the
adjustable setup variables:
Adaptive Filter =
On
PMT Temperature Set Point =
15°CNormal
Filter Size =
12 samples
Dwell Time =
7 seconds
Rate of Change(ROC) Threshold =
10%
Dynamic Span =
Off
Sample Time =
8 seconds
Reaction Cell Temperature =
50oC
Dynamic Zero =
Off
and with or without any of the following options:
180 Stainless Steel Valves 283 Internal Zero/Span With Valves (IZS) 356
Level One Spares Kit
184 Pump Pack 325 RS-232/Status Output 357 Level Two Spares Kit
280 Rack Mount With Slides 355 Expendables PE5 Permeation Tube for IZS
[Federal
Register:
Vol. 56, page 27014, 06/12/91]
Beckman Model 952-A NO/NO2/NOx
Analyzer
Automated Reference Method:
RFNA-0179-034
"Beckman Model 952-A NO/NO2/NOx
Analyzer," operated on the 0-0.5 ppm range with the 5-micron Teflon
sample filter (Beckman
P/N 861072 supplied with the analyzer) installed on the sample inlet
line, with or without the Remote Operation Option (Beckman No.
635539). [Federal
Register:
Vol. 44, page 7806, 02/07/79]
Bendix Model 8101-B Oxides of Nitrogen Analyzer
Automated Reference Method:
RFNA-0479-038
"Bendix Model 8101-B Oxides of Nitrogen Analyzer," operated on a 0-0.5
ppm range with a Teflon sample filter installed on the sample
inlet line and with the following post-manufacture modifications: 1)
Ozone generator and reaction chamber input-output tubing
modification per Bendix Service Bulletin 8101B-2; 2) The approved
converter material; 3) The revised and EPA-approved operation
and service manual. These items are mandatory and must be obtained from
ABB Process Analytics. The analyzer may be operated with
or without any of the following optional modifications: a. Perma Pure
dryer/ambient air modification; b. Valve cycle time modification;
c. Zero potentiometer centering modification per Bendix Service Bulletin
8101B-1; d. Reaction chamber vacuum gauge modification.
[Federal
Register:
Vol. 44, page 26792, 05/07/79]
Bendix/Combustion Engineering Model 8101-C Oxides of Nitrogen Analyzer
Automated Reference Method:
RFNA-0777-022
"Bendix or Combustion Engineering Model 8101-C Oxides of Nitrogen
Analyzer," operated on a 0-0.5 ppm range with a Teflon sample
filter (Bendix P/N 007163) installed on the sample inlet line. [Federal
Register:
Vol. 42, page 37435, 07/21/77]
Columbia Scientific Industries Models 1600 and 5600 Analyzers
Automated Reference Method:
RFNA-0977-025
"CSI Model 1600 Oxides of Nitrogen Analyzer," operated on a 0-0.5 ppm
range with a Teflon sample filter (CSI P/N M951-8023)
installed on the sample inlet line, with or without any of the following
options:
951-0103 Rack Ears 951-0114 Recorder Output, 5 V
951-0104 Rack Mounting Kit (Ears & Slides) 951-0115 External Pump (115
V, 60 Hz)
951-0106 Current Output, 4-20 mA (Non-Insulated) 951-8072 Molybdenum
Converter Assembly (Horizontal)
951-0108 Diagnostic Output Option 951-8074 Copper Converter Assembly
(Horizontal)
951-0111 Recorder Output, 10 V 951-8079 Copper Converter Assembly
(Vertical)
951-0112 Remote Zero/Span Sample Control 951-8085 Molybdenum Converter
Assembly (Vertical)
NOTE: The vertical molybdenum converter assembly is standard on all new
analyzers as of 1-1-87; however, use of any of the other
converter assemblies is optional. Also, the above options reflect new
CSI part numbers.
"CSI Model 5600 Oxides of Nitrogen Analyzer," operated on a 0-0.5 ppm
range, with any signal integration time in the range of 20 to
99 seconds, with a Teflon sample filter (CSI P/N M951-8023) installed on
the sample inlet line, and with or without any of the following
options:
954-0121 Status Contacts 964-0126 Printer 954-0131 Rack Mounting Kit
(ears and slides)
954-0122 Input Solenoids 954-8024 Cartridge Dryer 964-0012 Single Headed
Pump - Gast
954-0125 Current Output, 4-20 mA 951-0115 Single Headed Pump - KNF
[Federal
Register:
Vol. 42, page 46574, 09/16/77]
NO2
Dasibi Model 2108 Oxides of Nitrogen Analyzer
Automated Reference Method:
RFNA-1192-089
"Dasibi Model 2108 Oxides of Nitrogen Analyzer," operated on the 0-500
ppb range, with software revision 3.6 installed in the analyzer,
with the auto thumbwheel switch and the diag thumbwheel switch settings
at 0, with the following internal CPU dipswitch settings:switch
position function
1 open (down) Recorder outputs are NO & NO2
5 open (down) 3 minute time constant
6 closed (up) 3 minute time constant;
with a 5-micron Teflon filter element installed in the filter holder,
and with or without any of the following options:
Built-in Permeation Oven Rack Mounting Three-Channel Recorder Output
RS-232 Interface 4-20 mA Output [Federal
Register:
Vol. 57, page 55530, 11/25/92]
DKK-TOA Corporation Model GLN-114E Nitrogen Oxides Analyzer
Automated Reference Method:
RFNA-0798-121
“DKK-TOA Corporation Models GLN-114E and GLN-114E-1 Nitrogen Oxides
Analyzer,” operated within a temperature range of 20
to 30 degrees C, on any of the following measurement ranges: 0-0.0501,
0-0.1001,
0-0.2001,
0-0.500, and 0-1.000 ppm, and with or
without the optional Internal zero air supply and permeation tube oven.2
[Federal
Register:
Vol. 63, page 41253, 08/03/98]
Environnement S. A. Model AC31M NO2
Analyzer
Automated Reference Method:
RFNA-0795-104
"Environnement S. A. Model AC31M Chemiluminescent Nitrogen Oxide
Analyzer," operated with a full scale range of 0 - 500 ppb,
at any temperature in the range of 15°C to 35°C, with a 5-micron PTFE
sample particulate filter, with the following software settings:
Automatic response time ON; Minimum response time set to 60 seconds (RT
÷ 2); and with or without any of the following options:2
Internal Permeation Oven; Connection for Silica Gel Dryer; RS232-422
interface; EV3 valve; Internal Printer.
[Federal
Register:
Vol. 60, page 38326, 07/26/95]
Environnement S. A. Model AC32M NO2
Analyzer
Automated Reference Method:
RFNA-0202-146
“Environnement S. A. Model AC32M Chemiluminescent Nitrogen Oxides
Analyzer," operated with a full scale range of 0 - 500 ppb,
at any temperature in the range of 10°C to 35°C, with a 5-micron PTFE
sample particulate filter, with response time setting 11 (automatic
response time), and with or without the following option: Internal
permeation oven.
[Federal
Register:
Vol 67, page 15567, 04/02/02]
Environnement S.A. SANOA Multigas Longpath Monitoring System
Automated Reference Method:
EQNA-0400-139
“Environnement S.A. Model SANOA Multigas Longpath Air Quality Monitoring
System,” consisting of a receiver, one or more
projectors, interface unit, a user-provided control unit computer
running the SANOA VisionAIR software, and associated incidental
equipment; configured for measuring NO2,
with the temperature control and internal calibration cell options
installed, operated with a
measurement range of 0 to 0.5 ppm, over an installed monitoring path
length of between 27 and 500 meters, within an ambient air
temperature range of -30 to +45°C, with a measurement (integrating) time
of 180 seconds, and with or without external temperature and
barometric pressure sensors or any of the following options: external
(meteo) input connection, series 1M bus connection, OGR type
projector, analog outputs. [Federal
Register:
Vol 65, page 26603, 05/08/00]
Horiba Instruments Model APNA-360 NO-NO2-NOX
Monitor
Automated Reference Method:
RFNA-0196-111
"Horiba Instruments, Inc. Model APNA-360 Ambient NO-NO2-NOX
Monitor," operated with a full scale range of 0 - 0.50 or 0 - 1.0 ppm,
at any temperature in the range of 10 °C to 40 °C, with a Line Setting
of "MEASURE", and an Analog Output of "MOMENTARY
VALUE", and with or without the following options:2
1) Rack Mounting Plate and Side Rails 2) RS-232 Communications Port.
[Federal
Register:
Vol. 61, page 11404, 03/20/96]
Meloy Model NA530R Nitrogen Oxides Analyzer
Automated Reference Method:
RFNA-1078-031
"Meloy Model NA530R Nitrogen Oxides Analyzer," operated on the following
ranges and time constant switch positions:
Range, ppm: 0-0.11
0-0.251
0-0.5 0-1.0
Time Constant Setting: 4 3 or 4 2,3, or 4 2,3, or 4
Operation of the analyzer requires an external vacuum pump, either Meloy
Option N-10 or an equivalent pump capable of maintaining
a vacuum of 200 torr (22 inches mercury vacuum) or better at the pump
connection at the specified sample and ozone-air flow rates of
1200 and 200 cm3/min,
respectively. The analyzer may be operated at temperatures between 10°C
and 40°C and at line voltages between
105 and 130 volts, with or without any of the following options: N-1A
Automatic Zero And Span; N-2 Vacuum Gauge; N-4 Digital
Panel Meter; N-6 Remote Control For Zero And Span; N-6B Remote Zero/Span
Control And Status (Pulse); N-6C Remote Zero/Span
Control And Status (Timer); N-9 Manual Zero/Span; N-10 Vacuum Pump
Assembly (See Alternate Requirement Above); N-11 Auto
Ranging; N-14B Line Transmitter; N-18 Rack Mount Conversion; N-18A Rack
Mount Conversion.
[Federal
Register:
Vol. 43, page 50733, 10/31/78 and Vol. 44, page 8327, 02/09/79]
NO2
Monitor Labs Model 8440E Nitrogen Oxides Analyzer
Automated Reference Method:
RFNA-0677-021
"Monitor Labs Model 8440E Nitrogen Oxides Analyzer," operated on a 0-0.5
ppm range (position 2 of range switch) with a time constant
setting of 20 seconds, with or without any of the following options:
TF- Sample Particulate Filter DO- Status Outputs 018A- Ozone Dry Air
O18B- Ozone Dry Air - No Drierite
With TFE Filter Element R- Rack Mount V- Zero/Span Valves FM- Flow
meters
[Federal
Register:
Vol. 42, page 37434, 07/21/77; Vol. 42, page 46575, 09/16/77; Vol. 46,
page 29986, 06/04/81]
Monitor Labs/Lear Siegler Model 8840 Nitrogen Oxides Analyzer
Automated Reference Method:
RFNA-0280-042
"Monitor Labs or Lear Siegler Model 8840 Nitrogen Oxides Analyzer,"
operated on a range of either 0-0.5 or 0-1.0 ppm, with an internal
time constant setting of 60 seconds, a TFE sample filter installed on
the sample inlet line, with or without any of the following options:
02 Flowmeter 08A Pump Pac Assembly With 09A (115 VAC) 011A Recorder
Output 1 Volt
03A Rack Ears 08B Pump Pac Assembly With 09B (100 VAC) 011B Recorder
Output 100 mV
03B Slides 08C Pump Pac Assembly With 09C (220/240 VAC) 011C Recorder
Output 10 mV
05A Zero/Span Valves 08D Rack Mount Panel Assembly 012A DAS Output 1
Volt
05B Valve/Relay 09A Pump 115 VAC 50/60 Hz 012B DAS Output 100 mV
06 Status 09B Pump 100 VAC 50/60 Hz 012C DAS Output 10 mV
07A Input Power Transformer 100 VAC, 50/60 Hz 09C Pump 220/240 VAC 50 Hz
013A Ozone Dry Air
07B Input Power Transformer 220/240 VAC 50 Hz 013B Ozone Dry Air - No
Drierite
[Federal
Register:
Vol. 45, page 9100, 02/11/80 and Vol. 46, page 29986, 06/04/81]
Monitor Labs/Lear Siegler Model 8841 Nitrogen Oxides Analyzer
Automated Reference Method:
RFNA-0991-083
"Monitor Labs or Lear Siegler Model 8841 Nitrogen Oxides Analyzer,"
operated on the 0-0.05 ppm1,
0-0.1 ppm1,
0-0.2 ppm1,
0 - 0.5
ppm, or 0-1.0 ppm range, with manufacturer-supplied vacuum pump or
alternative user-supplied vacuum pump capable of providing
200 torr or better absolute vacuum while operating with the analyzer. [Federal
Register:
Vol. 56, page 47473, 9/19/91]
Opsis Model AR 500 and System 300 Open Path Ambient Air
Automated Equivalent Method:
EQNA-0495-102
Monitoring Systems for NO2
"Opsis Model AR 500 System" or "System 300" Open Path (long path)
Ambient Air Monitoring Systems, configured for measuring
NO2,
with one detector and movable grating, operated with a measurement range
of 0 to 0.5 ppm, an installed monitoring path length
between 50 and 500 meters (or 50 and 1000 meters with the ER 150 option,
AR 500 System only), xenon lamp type B (150 watt),
fiber optic cable length between 3 and 20 meters; operating within an
ambient air temperature range of -50 to +50°C, an analyzer
temperature range of 20 to 30°C, a measurement (integrating) time
setting between 30 and 120 seconds (0 min:30 sec. to 2 min:00
sec.), and with a complete cycle time of not more than 200 seconds (3
min, 20 sec.). Under this method designation, the Model AR
500 System or System 300 consists of: AR 500 opto-analyser; emitter EM
110 and receiver RE 110 (together identified as ER 110);
optic fibre cable OF60-S; power supply PS 150; OPSIS operational
software, version 7.0 or 7.1; and initial on-site installation,
setup, and limited operator training.2
Optional components that can be used with the Model AR 500 only,
in addition to or as alternative to corresponding components
listed above: • AR 503 opto-analyzer configured as Model AR 500 (only
the center detector active, sequential monitoring)
• Emitter/receiver ER 150 (for monitoring path lengths up to 1
kilometer) • Transceiver ER 130 and Retroreflector RE 090 with 7
prisms (max. monitoring path length 150 meters) or 12 prisms (max.
monitoring path length 250 meters) • Receiver RE 130
• Xenon lamp type A (higher short-wavelength UV output) • Optic fibre
cable OF60-R (low-loss for short wavelengths)
• Multiplexers MX 004 and MX 024 • Dataloggers DL 010 and DL 016 •
Analogue and digital input/output cards AO 008, AI 016,
and DI 032 • Analogue and digital isolation cards IA 008, ID 008, OA
008, and OD 008 • Window heaters HF 110 and HF 150
• Mirror heaters HM 110 and HM 150 • Auto calibration unit CU 007 •
Software packages IO 80 (for the analogue and digital
input/output adapters), DL10 and DL16 (for data loggers), ComVision, and
STAT 500;
Recommended calibration and accuracy audit components (or equivalent)
for either Model AR 500 or System 300:
• Wavelength calibration lamp CA 004 • Calibration bench CB 100 •
Receiver unit RE 060 (two required) • Calibration unit CA
150, with same type lamp as used in the monitoring path emitter • Power
supply PS 150 for calibration unit CA 150 • Calibration
cells CC 001-X, where X represents various cell lengths from 1 to 900 mm
• Filter GG 400 • Special calibration cells CC 110 or
CC 150 (for mounting directly on receiver) • Light meter LM 010 [Federal
Register:
Vol. 60, page 21518, 05/02/95].
Philips Model PW9762/02 NO/NO2/NOx
Analyzer
Automated Reference Method:
RFNA-0879-040
"Philips Model PW9762/02 NO/NO2/NOx
Analyzer," consisting of the following components: PW9762/02 Basic
Analyzer; PW9729/00
Converter Cartridge; PW9731/00 Sampler or PW9731/20 Dust Filter;
operated on a range of 0-0.5 ppm, with or without any of the
following accessories: PW9752/00 Air Sampler Manifold; PW9732/00 Sample
Line Heater; PW9011/00 Remote Control Set.
[Federal
Register:
Vol. 44, page 51683, 09/04/79]
NO2...LEAD
SIR S.A. Model S-5012 Nitrogen Oxides Analyzer
Automated Reference Method:
RFNA-0804-152
SIR S.A. Model S-5012 Chemiluminescent Nitrogen Oxides Analyzer,
operated with a full scale range of 0 - 500 ppb, at any temperature
in the range of 20°C to 30 °C, with the integration time set to 1
minute, with the “initial zero” disabled, and with a specified Teflon
particulate filter installed in the sample inlet filter holder.2
[Federal
Register:
Vol. 69, page 47924, 08/06/04]
Teledyne - Advanced Pollution Instrumentation, Inc. Models 200A, 200AU,
200E;
Automated Reference Method:
RFNA-1194-099
Teledyne Analytical Instruments Model 9110A; or Teledyne Monitor Labs
sensor-e™ Model TML-41 NO2
Analyzers
"Teledyne - Advanced Pollution Instrumentation, Inc. Models 200A, 200AU,
9110A, or 200E; Teledyne Analytical Instruments Model
9110A; or Teledyne Monitor Labs, Inc. sensor-e™ Model TML-41
Chemiluminescence Nitrogen Oxides Analyzer," operated on any
full scale range between 0-0.05 ppm and 0-1.0 ppm, with a PTFE filter
element installed in the internal filter assembly, with the following
software settings: dynamic zero: OFF or ON; dynamic span: OFF; cal-on-NO2:
OFF; dilution factor: OFF or set to 1.0; autocal: ON or
OFF; independent range: ON or OFF; autorange: ON or OFF;
temperature/pressure compensation: ON; and with or without any of the
following options (if available): rack mounts with or without slides,
rack mount for external pump, zero/span valves, 4-20 mA analog
outputs, status outputs, RS-232 output.
Models 200A, 200E2,
and TML-41 only:
operated at any temperature in the range of 5 °C to
40 °C, with either a user- or vendor-supplied vacuum pump capable of
providing an absolute pressure of 10 inches mercury or less at
2 slpm, with or without optional internal zero/span (IZS) and permeation
tubes for IZS, gold-plated reaction chamber, or Nafion-type
sample gas conditioner, ethernet output, control input, RS-485 output.
Model 200AU only:
operated at any temperature in the range
of 20 °C to 30 °C, with either a user- or vendor-supplied vacuum pump
capable of providing an absolute pressure of 4 inches mercury
or less at 1 slpm. [Federal
Register:
Vol. 59, page 61892,12/02/94]
Teledyne Monitor Labs/Ecotech Models ML9841 or ML9841A/EC9841A,
Automated Reference Method:
RFNA-1292-090
Teledyne Monitor Labs/Ecotech Model ML9841B/EC9841B, or Wedding &
Associates Model 1030 NO2
Analyzers
"Teledyne Monitor Labs Ecotech Models ML9841, ML9841A/EC9841A, or
ML9841B/EC9841B, or Wedding & Associates, Inc. Model
1030 Nitrogen Oxides Analyzers," operated on any full scale range
between 0-0.05 ppm1
and 0-1.0 ppm, at any temperature in the range
of 15°C to 35°C, with the service switch on the secondary panel set to
the
In
position; with the following menu choices selected: Range:
0.05 ppm
to
1.0 ppm;
Over-ranging:
Enabled
or
Disabled;
Calibration:
Manual
or
Timed;
Diagnostic Mode:
Operate;
Filter Type:
Kalman;
Pres/Temp/Flow Comp:
On;
Span Comp:
Disabled;
and as follows:
Models ML9841 and ML9841A/EC9841A
- with a fivemicron
Teflon® filter element installed internally, with the 50-pin I/O board
installed on the rear panel configured at any of the following
output range setting: Voltage, 0.1 V, 1 V, 5 V, 10 V; Current, 0-20 mA,
2-20 mA, 4-20 mA; and with or without any of the following
options: Valve Assembly for External Zero/Span (EZS); Internal Zero/Span
(IZS) Assembly for; Rack Mount Assembly; Internal Floppy
Disk Drive.
Models ML9841B/EC9841B and 1030
- with a vendor-supplied or equivalent user-supplied five-micron Teflon®
filter
and exhaust pump, and with or without any of the following options:
Valve Assembly for External Zero/Span (EZS); 50-pin I/O board;
Internal Zero/Span (IZS) Assembly; Rack Mount Assembly; Charcoal exhaust
scrubber; hinged, fold-down front panel.
[Federal
Register:
Vol. 57, page 60198, 12/18/92]
Thermo Electron/Thermo Environmental Instruments Model 14 B/E
Automated Reference Method:
RFNA-0179-035
"Thermo Electron or Thermo Environmental Instruments, Inc. Model 14 B/E
Chemiluminescent NO/NO2/NOx
Analyzer," operated on
the 0-0.5 ppm range, with or without any of the following options:
14-001 Teflon Particulate Filter 14-003 Long-Time Signal Integrator
14-005 Sample Flowmeter
14-002 Voltage Divider Card 14-004 Indicating Temperature Controller
14-006 Air Filter
[Federal
Register:
Vol. 44, page 7805, 02/07/79 and Vol.44, page 54545, 09/20/79]
Thermo Electron/Thermo Environmental Instruments Model 14 D/E
Automated Reference Method:
RFNA-0279-037
"Thermo Electron or Thermo Environmental Instruments, Inc. Model 14 D/E
Chemiluminescent NO/NO2/NOx
Analyzer," operated on
the 0-0.5 ppm range, with or without any of the following options:
14-001 Teflon Particulate Filter; 14-002 Voltage Divider Card.
[Federal
Register:
Vol. 44, page 10429, 02/20/79]
Thermo Environmental Instruments Models 42, 42C NO/NO2/NOx
Analyzer
Automated Reference Method:
RFNA-1289-074
"Thermo Environmental Instruments Inc. Model 42 or Model 42C NO-NO2-NOx
Analyzer," operated on any measurement range
between 0-50 ppb1
and 0-1000 ppb, with any time average setting from 10 to 300 seconds,
with temperature and/or pressure
compensation on or off, operated at temperatures between 15 °C and 35
°C, with or without an exhaust ozone scrubber or any of the
following options:
2
42-002 Rack mounts 42-006 Pressure transducer (Model 42 only)
42-003 Internal Zero/span and sample valves with remote activation
42-007 Ozone particulate filter
42-004 Sample/ozone flow meters (Model 42 only) 42-008 RS-232/485
interface
42-005 4-20 mA current output 42-009 Permeation dryer
[Federal
Register:
Vol. 54, page 50820, 12/11/89]
LEAD
LEAD
Reference Method for Lead
Manual Reference Method:
40 CFR Part 50, Appendix G
Reference Method for the Determination of Lead in Suspended Particulate
Matter Collected from Ambient Air.
[Federal
Register:
Vol. 43, page 46258, 10/05/78]
Energy-Dispersive X-Ray Fluorescence Spectrometry (TNRCC)
Manual Equivalent Method:
EQL-0783-058
"Determination of Lead Concentration in Ambient Particulate Matter by
Energy-Dispersive X-Ray Fluorescence Spectrometry (Texas
Natural Resource Conservation Commission)" Texas Natural Resource
Conservation Commission, P.O. Box 13087, Austin, TX 78711-
3087. [Federal
Register:
Vol. 48, page 29742, 06/28/83]
Energy-Dispersive X-Ray Fluorescence Spectrometry (NEA, Inc.)
Manual Equivalent Method:
EQL-0589-072
"Determination of Lead Concentration in Ambient Particulate Matter by
Energy-Dispersive X-Ray Fluorescence Spectrometry (NEA,
Inc.)" Nuclear Environmental Analysis, Inc., Suite 260, 10950 SW 5th
Street, Beaverton, OR 97005.
[Federal
Register:
Vol. 54, page 20193, 05/10/89]
Flame Atomic Absorption Spectrometry
Manual Equivalent Method:
EQL-0380-043
"Determination of Lead Concentration in Ambient Particulate Matter by
Flame Atomic Absorption Spectrometry Following Ultrasonic
Extraction with Heated HNO3-HCl"
Federal Register:
Vol. 45, page 14648, 03/06/80]
Flameless Atomic Absorption Spectrometry (EPA/RTP, N.C.)
Manual Equivalent Method:
EQL-0380-044
"Determination of Lead Concentration in Ambient Particulate Matter by
Flameless Atomic Absorption Spectrometry (EPA/RTP, N.C.)"
[Federal
Register:
Vol. 45, page 14648, 03/06/80]
Flameless (Graphite Furnace) Atomic Absorption (Houston, Texas)
Manual Equivalent Method:
EQL-0895-107
"Determination of Lead Concentration in Ambient Particulate Matter by
Flameless (Graphite Furnace) Atomic Absorption (City of
Houston, Texas)." Health and Human Services Department, Environmental
Chemistry Service, 1115 S. Braeswood, Houston, TX
77030. [Federal
Register:
Vol. 60, page 39383, 08/02/95]
Flameless Atomic Absorption Spectrometry (Omaha)
Manual Equivalent Method:
EQL-0785-059
"Determination of Lead Concentration in Ambient Particulate Matter by
Flameless Atomic Absorption Spectrometry (Omaha-Douglas
County Health Department)" Omaha-Douglas County Health Department, 1819
Farnam Street, Omaha, NE 68183.
[Federal
Register:
Vol. 50, page 37909, 09/18/85]
Inductively Coupled Argon Plasma-Optical Emission Spectrometry (Doe Run)
Manual Equivalent Method:
EQL-0196-113
"Determination of Lead Concentration in Ambient Particulate Matter by
Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (Doe Run Co.)" Doe Run Company, Smelting Division, 881 Main
Street Herculaneum, MO 63048
[Federal
Register:
Vol. 61, page 11404, 03/20/96]
Inductively Coupled Argon Plasma-Optical Emission Spectrometry (EPA/RTP)
Manual Equivalent Method:
EQL-0380-045
"Determination of Lead Concentration in Ambient Particulate Matter by
Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (EPA/RTP, N.C.)" [Federal
Register:
Vol. 45, page 14648, 03/06/80]
Inductively Coupled Argon Plasma-Optical Emission Spectrometry (IL)
Manual Equivalent Method:
EQL-1193-094
"Determination of Lead Concentration in Ambient Particulate Matter by
Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (State of Illinois)." State of Illinois, Environmental
Protection Agency, Champaign Inorganic Laboratory, 2120 South
First Street, Champaign, IL 61820 [Federal
Register:
Vol. 58, page 61902, 11/23/93]
Inductively Coupled Argon Plasma-Optical Emission Spectrometry (Kansas)
Manual Equivalent Method:
EQL-0592-085
"Determination of Lead Concentration in Ambient Particulate Matter by
Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (State of Kansas)" State of Kansas, Department of Health
and Environment, Forbes Field, Building 740, Topeka, KS
66620-0001. [Federal
Register:
Vol. 57, page 20823, 05/15/92]
Inductively Coupled Argon Plasma-Optical Emission Spectrometry (Montana)
Manual Equivalent Method:
EQL-0483-057
"Determination of Lead Concentration in Ambient Particulate Matter by
Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (State of Montana)". State of Montana, Department of Health
and Environmental Sciences, Cogswell Building, Helena,
MT 59620. [Federal
Register:
Vol. 48, page 14748, 04/05/83]
Inductively Coupled Argon Plasma-Optical Emission Spectrometry (NETI)
Manual Equivalent Method:
EQL-1188-069
"Determination of Lead Concentration in Ambient Particulate Matter by
Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (Northern Engineering and Testing, Inc.)" Northern
Engineering and Testing, Inc., P.O. Box 30615, Billings, MT 59107.
[Federal
Register:
Vol. 53, page 44947, 11/07/88]
Inductively Coupled Argon Plasma-Optical Emission Spectrometry (NH)
Manual Equivalent Method:
EQL-1290-080
"Determination of Lead Concentration in Ambient Particulate Matter by
Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (State of New Hampshire)" State of New Hampshire,
Department of Environmental Services, Laboratory Service Unit,
6 Hazen Drive (P.O. Box 95), Concord, NH 03302-0095. [Federal
Register:
Vol. 55, page 49119, 11/26/90]
Inductively Coupled Argon Plasma-Optical Emission Spectrometry (PA)
Manual Equivalent Method:
EQL-0592-086
"Determination of Lead Concentration in Ambient Particulate Matter by
Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (Commonwealth of Pennsylvania)" Commonwealth of
Pennsylvania, Department of Environmental Resources, P.O. Box
2357, Harrisburg, PA 17105-2357. [Federal
Register:
Vol. 57, page 20823, 05/15/92]
Inductively Coupled Argon Plasma-Optical Emission Spectrometry (Pima,
AZ)
Manual Equivalent Method:
EQL-0995-109
"Determination of Lead Concentration in Ambient Particulate Matter by
Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (Pima County, Arizona)." Pima County, Wastewater Management
Department, 201 North Stone Avenue, Tucson,
Arizona 85701-1207. [Federal Register: Vol. 60, page 54684,
10/25/95]
Inductively Coupled Argon Plasma-Mass Spectrometry (Pima Co., AZ)
Manual Equivalent Method:
EQL-0995-110
"Determination of Lead Concentration in Ambient Particulate Matter by
Inductively Coupled Plasma-Mass Spectrometry (Pima County,
Arizona)." Pima County, Wastewater Management Department, 201 North
Stone Avenue, Tucson, Arizona 85701-1207.
[Federal
Register:
Vol. 60, page 54684, 10/25/95]
Inductively Coupled Argon Plasma-Optical Emission Spectrometry (RI)
Manual Equivalent Method:
EQL-0888-068
"Determination of Lead Concentration in Ambient Particulate Matter by
Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (State of Rhode Island)," State of Rhode Island Department
of Health, Air Pollution Laboratory, 50 Orms Street,
Providence, RI 02904 [Federal
Register:
Vol. 53, page 30866, 08/16/88]
Inductively Coupled Argon Plasma-Optical Emission Spectrometry (Silver
Valley)
Manual Equivalent Method:
EQL-1288-070
"Determination of Lead Concentration in Ambient Particulate Matter by
Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (Silver Valley Laboratories)," Silver Valley Laboratories,
Inc., P.O. Box 929, Kellogg, ID 83837.
[Federal
Register:
Vol. 53, page 48974, 12/05/88]
Inductively Coupled Argon Plasma-Atomic Emission Spectrometry (TNRCC)
Manual Equivalent Method:
EQL-0400-140
"Determination of Lead Concentration in Ambient Particulate Matter by
Inductively Coupled Argon Plasma-Atomic Emission
Spectrometry (TNRCC),” Texas Natural Resource Conservation Commission
Laboratory, 5144 E. Sam Houston Parkway N., Houston,
TX 77030. [Federal
Register:
Vol 65, page 26603, 5/8/00]
Inductively Coupled Argon Plasma-Optical Emission Spectrometry (WV)
Manual Equivalent Method:
EQL-0694-096
"Determination of Lead Concentration in Ambient Particulate Matter by
Inductively Coupled Argon Plasma-Optical Emission
Spectrometry (State of West Virginia)." State of West Virginia,
Department of Commerce, Labor and Environmental Resources,
Division of Environmental Protection, 1558 Washington Street East,
Charleston, WV 25311-2599
[Federal
Register:
Vol. 59, page 29429, 06/07/94]
Wavelength Dispersive X-Ray Fluorescence Spectrometry (CA)
Manual Equivalent Method:
EQL-0581-052
"Determination of Lead Concentration in Ambient Particulate Matter by
Wavelength Dispersive X-Ray Fluorescence Spectrometry"
California Department of Health Services, Air & Industrial Hygiene
Laboratory, 2151 Berkeley Way, Berkeley, CA 94704.
[Federal
Register:
Vol. 46, page 29986, 06/04/81]
NOTES
1
Users should be aware that designation of this analyzer for operation on
ranges less than the range specified in the performance
specifications for this analyzer (40 CFR 53, Subpart B) is based on
meeting the same absolute performance specifications required for the
specified range. Thus, designation of these lower ranges does not imply
commensurably better performance than that obtained on the
specified range.
2
This analyzer is approved for use, with proper factory configuration, on
either 50 or 60 Hertz line frequency and nominal power line
voltages of 115 Vac and 230 Vac.
Sources or Contacts for Designated Reference and Equivalent Methods
ABB Process Analytics
P.O. Box 831
Lewisburg, WV 24901
(304) 647-4358
Advanced Pollution Instrumentation, Inc.
[Refer to Teledyne - Advanced Pollution
Instrumentation, Inc.]
Andersen Instruments
[Refer to Thermo Electron Corporation]
www.thermoandersen.com
ASARCO Incorporated
3422 South 700 West
Salt Lake City, UT 84119
(801) 262-2459
Beckman Instruments, Inc.
Process Instruments Division
2500 Harbor Blvd.
Fullerton, CA 92634
(714) 871-4848
Bendix
[Refer to ABB Process Analytics]
BGI Incorporated
58 Guinan Street
Waltham, MA 02451
(781) 891-9380
www.bgiusa.com
(bgiinc@attglobal.net)
Columbia Scientific Industries
11950 Jollyville Road
Austin, TX 78759
(800) 531-5003
Combustion Engineering
[Refer to ABB Process Analytics]
Dasibi Environmental Corp.
[506 Paula Avenue
Glendale, CA 91201
(818) 247-7601
www.dasibi.com
]
DKK-TOA Corporation
29-10, 1-Chome, Takadanobaba,
Shinjuku-ku
Tokyo 169-8648, Japan
www.toadkk.co.jp
Ecotech Pty. Ltd.
12 Apollo Court
Blackburn, Victoria, 3130, Australia
+61 3 9894 2399
www.ecotech.com.au
Environnement S.A
111, bd Robespierre
78300 Poissy, France
www.environnement-sa.com
Instruments also available from:
Altech/Environnement U.S.A.
2623 Kaneville Court
Geneva, IL 60134
(630) 262- 4400
Environics, Inc.
69 Industrial Park Rd. E.
Tolland, CT 06084-2805
(203) 429-0077
www.environics.com
Graseby GMW
[Refer to Thermo Electron Corporation]
www.thermoandersen.com
Horiba Instruments Incorporated
17671 Armstrong Avenue
Irvine, CA 92714
(800) 446-7422
www.horiba.com
Lear Siegler
[Refer to Teledyne Monitor Labs, Inc.]
Commonwealth of Massachusetts
Department of Environmental
Quality Engineering
Tewksbury, MA 01876
Met One Instruments, Inc.
1600 Washington Blvd.
Grants Pass, OR 97526
(541) 471-7111
www.metone.com
(metone@metone.com)
McMillan
[Refer to Columbia Scientific Industries]
Mine Safety Appliances
600 Penn Center Blvd.
Pittsburgh, PA 15235-5810
(412) 273-5101
Monitor Labs, Inc.
[Refer to Teledyne Monitor Labs, Inc.]
New Star Environmental, LLC
3293 Ashburton Chase NE
Roswell, GA 30075
(770) 998-2590
Opsis AB, Furulund, Sweden
Instruments also available from:
Opsis, Inc.
146-148 Sound Beach Avenue
Old Greenwich, CT 06870
(203) 698-1810
www.opsis.se
State of Oregon
Department of Environmental Quality
Air Quality Division
811 S.W. Sixth Avenue
Portland, OR 97204
PCI Ozone Corp.
One Fairfield Crescent
West Caldwell, NJ 07006
(201) 575-7052
www.pci-wedeco.com
Phillips Electronic Instruments, Inc.
85 McKee Drive
Mahwah, NJ 07430
Rupprecht & Patashnick Co., Inc.
26 Tech Valley Drive
East Greenbush, NY 12061
(518) 452-0065
www.rpco.com
Sibata Scientific Technology, Ltd.
1-25, 3-chome
Ikenohata, Taito-ku
Tokyo 110, Japan
81-3(3822)2272
TTani@email.msn.com
SIR S.A. (Sistemas Instalaciones y Redes, S.A.)
Avenida de la Industria, 3
Tres Cantos 28760 Madrid, SPAIN
(+34) 91 803 66 02
http://www.sirsa.es
Teledyne - Advanced Pollution
Instrumentation, Inc.
6565 Nancy Ridge Drive
San Diego, CA 92121-2251
(619) 657-9800
www.teledyne-api.com
Teledyne Analytical Instruments
16830 Chestnut Street
City of Industry, CA 91748
(626) 934-1622
Teledyne Monitor Labs, Inc.
74 Inverness Drive East
Englewood, CO 80112-5189
(303) 792-3300
www.teledyne-ml.com
Thermo Electron Corporation
27 Forge Parkway
Franklin, MA 02038
(508) 520-0430 / (866) 282-0430
www.thermo.com
Thermo Environmental Instruments, Inc.
[Refer to Thermo Electron Corporation]
www.thermoei.com
Tisch Environmental, Inc.
145 S. Miami Avenue
Village of Cleves, OH 45002
(513) 467-9000
www.tisch-env.com
URG Corporation
116 Merritt Mill Road
Chapel Hill, NC 27516
(919) 942-2753
U.S. EPA
National Exposure Research Laboratory
Human Exposure & Atmospheric
Sciences Division
Process Modeling Research Branch (MD D205-
03)
Research Triangle Park, NC 27711
(919) 541- 3737
www.epa.gov/heasd
Wedding and Associates, Inc.
[Refer to Thermo Electron Corporation.]
U.S.
EPA REFERENCE & EQUIVALENT METHODS FOR AMBIENT AIR
Designation Method Designation Method
Method Number Code Method Number Code
SO2
Manual Methods
Reference method (pararosaniline) -- 097
Technicon I (pararosaniline) EQS-0775-001 097
Technicon II (pararosaniline) EQS-0775-002 097
SO2
Analyzers
Advanced Pollution Instr. 100 EQSA-0990-077 077
Advanced Pollution Instr. 100A/100AS EQSA-0495-100 100
Asarco 500 EQSA-0877-024 024
Beckman 953 EQSA-0678-029 029
Bendix 8303 EQSA-1078-030 030
Columbia Scientific Industries 5700 EQSA-0494-095 095
Dasibi 4108 EQSA-1086-061 061
DKK-TOA Corp. GFS-32 EQSA-0701-115 115
DKK-TOA Corp. GFS-112E, GFS-112E-1 EQSA-0100-133 133
Ecotech ML9850/EC9850, ML9850B/EC9850B EQSA-0193-092 092
Environnement S.A AF21M EQSA-0292-084 084
Environnement S.A AF22M EQSA-0802-149 149
Environnement S.A. SANOA EQSA-0400-138 138
Horiba Model APSA-360/APSA-360ACE EQSA-0197-114 114
Lear Siegler AM2020 EQSA-1280-049 049
Lear Siegler SM1000 EQSA-1275-005 005
Meloy SA185-2A EQSA-1275-006 006
Meloy SA285E EQSA-1078-032 032
Meloy SA700 EQSA-0580-046 046
Monitor Labs 8450 EQSA-0876-013 513
Monitor Labs or Lear Siegler 8850 EQSA-0779-039 039
Monitor Labs or Lear Siegler 8850S EQSA-0390-075 075
Opsis AR 500, System 300 (open path) EQSA-0495-101 101
Philips PW9700 EQSA-0876-011 511
Philips PW9755 EQSA-0676-010 010
Teledyne-Advanced Pollution Inst. 100E EQSA-0495-100 100
Teledyne Analytical Instruments 6400A EQSA-0495-100 100
Teledyne Monitor Labs ML9850, ML9850B EQSA-0193-092 092
Teledyne Monitor Labs TML-50 EQSA-0495-100 100
Thermo Electron 43 EQSA-0276-009 009
Thermo Electron 43A or Thermo
Environmental Instruments 43B, 43C EQSA-0486-060 060
Wedding 1040 EQSA-0193-092 092
O3
Analyzers
Advanced Pollution Instr. 400/400A/400E EQOA-0992-087 087
Beckman 950A RFOA-0577-020 020
Bendix 8002 RFOA-0176-007 007
Columbia Scientific Industries 2000 RFOA-0279-036 036
Dasibi 1003-AH, -PC, -RS EQOA-0577-019 019
Dasibi 1008-AH, -PC, -RS EQOA-0383-056 056
DKK-TOA Corp. GUX-113E, GUX-113E-1 EQOA-0200-134 134
Ecotech ML9810/EC9810, -9810B, -9811, -9812 EQOA-0193-091 091
Environics 300 EQOA-0990-078 078
Environnement S.A O341M EQOA-0895-105 105
Environnement S.A O342M EQOA-0206-148 148
Environnement S.A SANOA EQOA-0400-137 137
Horiba APOA-360 EQOA-0196-112 112
McMillan 1100-1 RFOA-1076-014 514
McMillan 1100-2 RFOA-1076-015 515
McMillan 1100-3 RFOA-1076-016 016
Meloy OA325-2R RFOA-1075-003 003
Meloy OA350-2R RFOA-1075-004 004
Monitor Labs 8410E RFOA-1176-017 017
Monitor Labs or Lear Siegler 8810 EQOA-0881-053 053
Opsis AR 500, System 300 (open path) EQOA-0495-103 103
PCI Ozone Corp. LC-12 EQOA-0382-055 055
Philips PW9771 EQOA-0777-023 023
Teledyne - Advanced Pollution Instr. 400E EQOA-0992-087 087
Teledyne Monitor Labs ML9810/9810B,
ML9811, ML9812 EQOA-0193-091 091
Teledyne Monitor Labs TML-10 EQOA-0992-087 087
Thermo Electron or Thermo
Environmental Instruments 49, 49C EQOA-0880-047 047
Wedding 1010 EQOA-0193-091 091
CO Analyzers
Beckman 866 RFCA-0876-012 012
Bendix 8501-5CA RFCA-0276-008 008
Dasibi 3003 RFCA-0381-051 051
Dasibi 3008 RFCA-0488-067 067
Ecotech ML9830/EC9830, ML9830B/EC9830B RFCA-0992-088 088
Environnement S.A CO11M RFCA-0995-108 108
Environnement S.A CO12M FRCA-0206-147 147
Horiba AQM-10, -11, -12 RFCA-1278-033 033
Horiba 300E/300SE RFCA-1180-048 048
Horiba APMA-360 RFCA-0895-106 106
MASS - CO 1 (Massachusetts) RFCA-1280-050 050
Monitor Labs 8310 RFCA-0979-041 041
Monitor Labs or Lear Siegler 8830 RFCA-0388-066 066
MSA 202S RFCA-0177-018 018
Teledyne Advanced Pollution Instr. 300 or 300E RFCA-1093-093 093
Teledyne Monitor Labs ML9830/9830B, RFCA-0992-088 088
Teledyne Monitor Labs TML-30 RFCA-1093-093 093
Thermo Electron or Thermo
Environmental Instruments 48, 48C RFCA-0981-054 054
Wedding 1020 RFCA-0992-088 088
NO2
Manual Methods
Sodium arsenite (orifice) EQN-1277-026 084
Sodium arsenite/Technicon II EQN-1277-027 084
TGS-ANSA (orifice) EQN-1277-028 098
NO2
Analyzers
Advanced Pollution Instr. 200 RFNA-0691-082 082
Advanced Pollution Instr. 200A/200AU RFNA-1194-099 099
Beckman 952A RFNA-0179-034 034
Bendix 8101-B RFNA-0479-038 038
Bendix 8101-C RFNA-0777-022 022
Columbia Scientific Indust.1600, 5600 RFNA-0977-025 025
Dasibi 2108 RFNA-1192-089 089
DKK-TOA Corp GLN-114E, GLN-114E-1 RFNA-0798-121 121
Ecotech ML9841A/EC9841A,ML9841B/EC9841B RFNA-1292-090 090
Environnement S.A. AC31M RFNA-0795-104 104
Environnement S.A. AC32M RFNA-0202-146 146
Environnement S.A. SANOA EQNA-0400-139 139
Horiba APNA-360 RFNA-0196-111 111
Meloy NA530R RFNA-1078-031 031
Monitor Labs 8440E RFNA-0677-021 021
Monitor Labs or Lear Siegler 8840 RFNA-0280-042 042
Monitor Labs or Lear Siegler 8841 RFNA-0991-083 083
Monitor Labs ML9841 RFNA-1292-090 090
Opsis AR 500, System 300 (open path) EQNA-0495-102 102
Philips PW9762/02 RFNA-0879-040 040
SIR S-5012 RFNA-0804-152 152
Teledyne-Advanced Pollution Inst. 200E RFNA-1194-099 099
Teledyne Analytical Instruments 9110A RFNA-1194-099 099
Teledyne Monitor Labs ML9841, ML9841A,
ML9841B RFNA-1292-090 090
Teledyne Monitor Labs TML-41 RFNA-1194-099 099
Thermo Electron or Thermo
Environmental Instruments 14B/E RFNA-0179-035 035
Thermo Electron or Thermo
Environmental Instruments 14D/E RFNA-0279-037 037
Thermo Environmental Instr. 42, 42C RFNA-1289-074 074
Wedding 1030 RFNA-1292-090 090
Pb Manual Methods
Reference method (hi-vol/AA spect.) -- 803
Hi-vol/AA spect. (alt. extr.) EQL-0380-043 043
Hi-vol/Energy-disp XRF (TX ACB) EQL-0783-058 058
Hi-vol/Energy-disp XRF (NEA) EQL-0589-072 072
Hi-vol/Flameless AA (EMSL/EPA) EQL-0380-044 044
Hi-vol/Flameless AA (Houston) EQL-0895-107 107
Hi-vol/Flameless AA (Omaha) EQL-0785-059 059
Hi-vol/ICAP spect. (Doe Run Co.) EQL-0196-113 113
Hi-vol/ICAP spect. (EMSL/EPA) EQL-0380-045 045
Hi-vol/ICAP spect. (Illinois) EQL-1193-094 094
Hi-vol/ICAP spect. (Kansas) EQL-0592-085 085
Hi-vol/ICAP spect. (Montana) EQL-0483-057 057
Hi-vol/ICAP spect. (NE&T) EQL-1188-069 069
Hi-vol/ICAP spect. (New Hampshire) EQL-1290-080 080
Hi-vol/ICAP spect. (Pennsylvania) EQL-0592-086 086
Hi-vol/ICAP spect. (Pima Co.,AZ) EQL-0995-109 109
Hi-vol/ICAP spect. (Pima Co.,AZ) EQL-0995-110 110
Hi-vol/ICAP spect. (Rhode Island) EQL-0888-068 068
Pb Manual Methods (cont’d)
Hi-vol/ICAP spect. (Silver Val. Labs) EQL-1288-070 070
Hi-vol/ICAP spect. (TNRCC) EQL-0400-140 140
Hi-vol/ICAP spect. (West Virginia) EQL-0694-096 096
Hi-vol/WL-disp. XRF (CA A&IHL) EQL-0581-052 052
PM10
Samplers
U.S.
EPA REFERENCE & EQUIVALENT METHODS FOR AMBIENT AIR
Designation Method Designation Method
Method Number Code Method Number Code
Andersen Instruments RAAS10-100 RFPS-0699-130 130
Andersen Instruments RAAS10-200 RFPS-0699-131 131
Andersen Instruments RAAS10-300 RFPS-0699-132 132
BGI Model PQ100 RFPS-1298-124 124
BGI Model PQ200 RFPS-1298-125 125
New Star Environmental Model NS-6070 RFPS-0202-141 141
Oregon DEQ Medium volume sampler RFPS-0389-071 071
Rupprecht & Patashnick Partisol 2000 RFPS-0694-098 098
R & P Partisol-FRM Model 2000 RFPS-1298-126 126
R & P Partisol-Plus Model 2025 Seq. RFPS-1298-127 127
Sierra-Andersen/GMW 1200 RFPS-1287-063 063
Sierra-Andersen/GMW 321-B RFPS-1287-064 064
Sierra-Andersen/GMW 321-C RFPS-1287-065 065
Sierra-Andersen/GMW 241 Dichot. RFPS-0789-073 073
Tisch Environmental Model TE-6070 RFPS-0202-141 141
W&A/Thermo Electron Mod 600 HVL RFPS-1087-062 062
PM10
Analyzers
Andersen Instruments FH62I-N Beta EQPM-0990-076 076
Environnement S.A. MP101M Beta EQPM-0404-151 151
Met One BAM1020, GBAM1020,
BAM1020-1, GBAM1020-1 Beta EQPM-0798-122 122
R & P TEOM 1400, 1400a EQPM-1090-079 079
Thermo Andersen Series FH 62 C14 Beta Monitor EQPM-1102-150 150
W&A/Thermo Electron 650 Beta Gauge EQPM-0391-081 081
PM2.5
Samplers
Andersen Model RAAS2.5-200 Audit RFPS-0299-128 128
BGI PQ200/200A RFPS-0498-116 116
BGI PQ200-VSCC or PQ200A-VSCC EQPM-0202-142 142
Graseby Andersen RAAS2.5-100 RFPS-0598-119 119
Graseby Andersen RAAS2.5-300 RFPS-0598-120 120
R & P Partisol-FRM 2000 PM-2.5 RFPS-0498-117 117
R & P Partisol-FRM 2000 PM-2.5 FEM EQPM-0202-143 143
R & P Partisol 2000 PM-2.5 Audit RFPS-0499-129 129
R & P Partisol 2000 PM-2.5 FEM Audit EQPM-0202-144 144
R & P Partisol-Plus 2025 PM-2.5 Seq. RFPS-0498-118 118
R & P Partisol-Plus 2025 PM-2.5 FEM Seq. EQPM-0202-145 145
Thermo Electron RAAS2.5-100 FEM EQPM-0804-153 153
Thermo Electron RAAS2.5-200 FEM EQPM-0804-154 154
Thermo Electron RAAS2.5-300 FEM EQPM-0804-155 155
Thermo Environmental Model 605 CAPS RFPS-1098-123 123
URG-MASS100 RFPS-0400-135 135
URG-MASS300 RFPS-0400-136 136
TSP Manual Method
Reference method (high-volume) -- 802 |