ANALIZADORES CONTINUOS
  Total Hydrocarbon Analyzer - Model 51C
This total hydrocarbon analyzer offers flexibility and reliability by using FID

This dedicated hydrocarbon analyzer is a microprocessor-based flame ionization detector which meets EPA Method 25A Compliance Monitoring criteria. An integral temperature-controlled heated zone eliminates condensation problems.

Product Details

The Model 51C Total Hydrocarbon Analyzer offers both flexibility and reliability by combining proven Flame Ionization Detection (FID) technology with ease to use menu-driven software and automatic diagnostics. Advanced eletronics provide unsurpassed stability and analytical performance.

The Model 51C Total Hydrocarbon Analyzer utilizes an easily understood and readable multi-line alphanumeric display. The advantages of an alphanumeric display extend to remote RS-232 operation using TEI for Windows software, while the same RS-232 communication protocols can be used with other data acquisition systems for added flexibility. Areas of application include incinerator emission, VOC abatement and control systems, process control, and other source monitoring applications.



15C
Key Features

• Four programmable analog outputs
• Internal zero and span values standard
• Low maintenance FID design
• Self Optimizing fuel / air mix
• Externally accessible inlet filter
• Four independant temperature control zones
• Low vibration pumps and noise damping


Specifications

Ranges 0-10,000 ppmv auto-ranging, display 0-5 to 0-10,000 ppm analog output user selectable
Minimum Detectable Limit 0.05 ppmv carbon
Drift ‹ 1% per 24 hours
Response Time ‹ 5 seconds for 90% of full scale
Repeatability 2% of reading
Accuracy 2% of measured value from 10% to 120% of span value
Sample Flow Rate 1.0 LPM nominal
Combustion Air 200-300 milimeters HC free air
Fuel 0~25 milimeters ultra pure H2 or 75 milimeters H2/He Mixture
Power Requirements 90-100 VAC @ 50/60 Hz, 105-125 VAC @ 50/60Hz, 210-250 VAC @ 50/60Hz
Dimensions 16.75"(W) x 8.62"(H) x 23"(D)
Weight 63 lbs.
Outputs voltage, current, RS-232/485
 

 
Dock Norte Argentina:
Cochabamba 477. Buenos Aires (1150AAE). Argentina. Tel: 54.11.4.307.8363 - 54.11.4.361.5807